Error Code
Error Buffer Text
Description of Error
Possible Problem
003
A03: Meter data CRC
Invalid calibration data
Meter not calibrated or data corrupt
040
A40: Probe read
Probe EPROM read error
Probe EPROM missing or defective
043
A43: Probe CRC 1
Probe CRC failed
Probe EPROM defective
044
A44: Probe hi current
Incorrect probe current
Probe EPROM error or meter malfunction
045
A45: Probe CRC 2
Probe CRC failed
Probe EPROM not programmed
051
A51: Probe status cmd CRC Probe status cmd CRC error
Probe EPROM defective
052
A52: Probe status msg CRC Probe status msg CRC error
Probe EPROM defective
053
A53: Probe page cmd CRC
Probe page cmd CRC error
Probe EPROM defective
054
A54: Probe page data CRC
Probe page data CRC error
Probe EPROM defective
098
A98: Probe data timeout
Probe data timeout
Internal meter error
099
A99: Cal data timeout
Calibration data timeout
Internal meter messaging / reset problem
060
D60: A/D
ADC error
Internal meter problem
061
D61: A/D zero
AD zero error
Internal meter problem
062
D62: Chip comm
PIC-DSP com error
Internal meter problem
063
D63: Offset D/A
DAC offset error
Internal meter problem
064
D64: I hall D/A
Probe current error
Broken probe or meter problem
065
D65: Output D/A
Output DAC error
Internal meter problem
066
D66: Gain
Amplifier gain error
Internal meter problem
067
D67: High input res
High input resistance
Broken probe or meter problem
Most errors result from a damaged probe. If an error code related to a probe problem appears, substitute another
probe if one is available to help determine if the probe may be defective.
MODEL 5200 ERROR CODES THAT MAY BE DISPLAYED IF THERE IS A PROBLEM
4-25