3-15
HOLD MODE SELECTION
To select Min, Max, Peak, Off, press and hold on the home screen
———————————————————————————
Press Range until the desired range is selected and shown in the info box
——————————————————
Alternatively, press
————————————————————————————————————————————————
To navigate to setting page 1, press
—————————————————————————————————
or
To toggle Range modes, press
———————————————————————————————
To return to normal operation, press
—————————————————————————————————————
To reset the currently held value, press on the home screen
——————————————————————————————
In some applications it may be desirable to hold a reading that is either greater than or less than all previous readings, or which has the
greatest peak value whether positive or negative.
The MAX HOLD function holds the reading that is arithmetically greater than all previous readings. For instance, a reading of +125.0 is greater
than +99.0 or -150.0.
The MIN HOLD function holds the reading that is arithmetically less than all previous readings. For instance, a reading of
-125.0 is less than 99.0 or +150.0.
The PEAK HOLD function captures and holds the peak value of the flux density waveform within the response time capabilities of the meter.
See the
SPECIFICATIONS
section of this manual for more information. The peak can be either positive or negative, whichever has the
greatest magnitude. For instance, a peak value of -100.0 is greater than a peak value of +90.0.
Do not use the AUTO RANGE when using HOLD. Use one of the available ranges are listed in the
SPECIFICATIONS
section of this manual.
Difference between PEAK and MAX for AC: Max is the maximum RMS value. PEAK gives the value of the point on the waveform with the
greatest amplitude.