POWER-UP
Power ON
To turn the unit ON, press
—————————————————————————————————————————————
Power OFF
To turn the unit OFF, for three seconds press
————————————————————————————————————
Alternatively, short press
—————————————————————————————————————————————
then select
————————————————————————————————————————————————
To cancel the power off operation, select
——————————————————————————————————————
When battery power is used for the first time, momentarily apply USB power to the meter to activate the battery. This will need to be done any
time the battery is disconnected and re-installed.
Press the POWER switch. There will be a momentary audible beep (If audio is enabled).
The instrument will conduct a self test before measurements begin. If a problem is detected the phrase “Err” will appear on the display followed
by a 3 digit code. The circuitry that failed will be retested and the error code will appear after each failure. This process will continue indefinitely
or until the circuitry passes the test. A condition in which a circuit fails and then passes should not be ignored because it indicates an
intermittent problem that should be corrected.
If the self test is successful, the meter will perform a self calibration. During this phase, the meter will display a calibration sequence "CALX"
message. Calibration will not occur if there is no probe connected. Until the probe is connected the phrase “No probe connected” will appear in
the box where readings are normally displayed.
Per the
SPECIFICATIONS
section, allow 15 minutes warmup time for rated accuracy.
The most common errors displayed are a result of damage to the sensing element located in the tip of the probe. A damaged sensing element
can result in the following error codes: E064 and E067.
Figure 3-5
Missing Probe Indication
Figure 3-6
Power-Up Display
3-5