    Page 26 of 56 

 

Manual Cleaning of the Reusable System 

1.

 

Prepare a neutral pH or instrument-safe enzymatic cleaning solution according to the cleaning solution 
manufacturer specifications. 

 

2.

 

Using the lock nut wrench that is provided with the system, remove the locking knob and locking nut. 

 

NOTE: 

This is a left-handed thread

Turning the knob and nut clockwise loosens or removes them. 

NOTE: 

These are two separate components so keep them together with other components and take care not to 

misplace them. 

 

 

 

Figure 2 

 

 

Immerse

 

 

Do Not Immerse

 

3.

 

With the exception of the hand piece and 
pneumatic hose, immerse the components 
into the cleaning solution and let soak for a 
minimum of two minutes. Do not

 

immerse 

the unit overnight.  

 
NOTE: 

For additional part number 

information, refer to Appendix B: Parts 
Information.

  

 

Figure 3 

 

 

Figure 4 

4.

 

Using a lint-free wipe moistened with the 
cleaning solution, wipe the hand piece and 
the pneumatic hose for a minimum of one 
minute each.  

 
CAUTION: 

Do not immerse the hand piece 

or pneumatic hose into the cleaning 
solution. 

 

 

Содержание Amalgatome SD

Страница 1: ...Amalgatome SD User Guide EXSURCO MEDICAL 10804 GREEN ROAD WAKEMAN OH 44889 PN X101456 REV C...

Страница 2: ...This page intentionally left blank...

Страница 3: ...malgatome SD System It is designed for use by qualified and trained medical professionals where the patient is under supervision of trained personnel Review this user guide paying particular attention...

Страница 4: ...This page intentionally left blank...

Страница 5: ...Warnings 9 Precautions 9 Setup Instructions 11 Inspection 11 Setup 11 System Check 15 Depth Gauge Setting 16 Operating Instructions 17 Graft Harvesting and Debridement 17 Postoperative Procedure 20 Cl...

Страница 6: ...ppendix A Technical Specifications 42 Specifications 42 Appendix B Parts Information 44 Ordering Information 46 Appendix C Glossary of Symbols 48 Appendix D Quick Reference Guides 50 Setup 50 Postoper...

Страница 7: ...partnerships and deep customer understanding We are committed to improving patients lives and outcomes by advancing the healing power of skin through innovation product development and the marketing o...

Страница 8: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 4 of 56 This page intentionally left blank...

Страница 9: ...rvesting Description The Amalgatome SD System is a pneumatically powered surgical skin grafting instrument capable of producing varying thicknesses of skin grafts from 0 005 to 0 045 inches in increme...

Страница 10: ...ring blade head 4 2 inch Cutting Head excision ring blade head 5 2 inch Lock Ring 6 Hand Piece Drive Unit 7 Lock Nut Wrench 8 Lock Ring Removal Tool Disposable Excision ring blades 2 inch and 4 inch...

Страница 11: ...ontains the 4 inch cutting head o SDSYSTEM02 contains the 2 inch cutting head The system is provided non sterile and is reusable after cleaning and sterilization prior to use according to the procedur...

Страница 12: ...s not recommended for use on the following locations on the body or where surgeon determines due to health reasons Eyelid s Post auricular region behind the ear and on the ear Inguinal region groin Ge...

Страница 13: ...sult from use of other manufacturer s excision ring blades or reuse of the Amalgatome SD excision ring blades Dispose of the excision ring blades in accordance with facility protocol for sharps contai...

Страница 14: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 10 of 56 This page intentionally left blank...

Страница 15: ...s in order to ensure that the parts function through the intended range of motion Do not use or reprocess damaged or defective parts If the system does not function correctly contact Exsurco Medical C...

Страница 16: ...d with peel corners pull open the package allowing the sterile operator to reach in and grab the clamshell WARNING SHARP NOTE It is best practice to have the package opened for the sterile operator It...

Страница 17: ...r seating Figure 2g g Place the excision ring blade lock ring over the excision ring blade and twist it in a clockwise motion until it stops Do not over tighten 3 Attach the pneumatic hose Figure 3a F...

Страница 18: ...pneumatic hose to the sterile field taking care not to compress the hose Figure 3e e Connect the male end of the pneumatic hose to the appropriate air source in the operating room designated for the...

Страница 19: ...xcision ring blade spins 3 Verify that all connections are secure and that the air pressure is set correctly while the system is running NOTE The required air pressure setting is 100 psi Figure 4 4 Re...

Страница 20: ...he individual patient demographics e g age weight ethnicity etc Rotate the depth adjustment knob to the left counterclockwise for thinner settings and to the right clockwise for thicker settings The d...

Страница 21: ...the depth gauge as necessary to the optimal setting The surgical site should be selected per facility protocol for the surgical procedure as determined for skin grafting or wound debridement Skin shou...

Страница 22: ...to emerge from the cutting head 6 Use either of the following methods to collect the recovered graft or debrided tissue Method 1 Use tissue forceps to gently lift the graft as it emerges from the cutt...

Страница 23: ...on Figure 6c Cutting portion or trailing edge of the instrument held parallel to the patient Demonstrating good excision ring blade engagement Figure 6d Cutting portion or trailing edge of the instrum...

Страница 24: ...ft has been recovered or the wound debridement procedure is completed and the Amalgatome SD has been returned to the surgical back table or instrument table Refer to Appendix D Quick Reference Guides...

Страница 25: ...aning agents with chlorine or chloride e g saline or bleach as the active ingredient as they are corrosive to stainless steel and can result in physical damage to the system including rust and corrosi...

Страница 26: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 22 of 56 This page intentionally left blank...

Страница 27: ...for hand washing use of personal protective equipment PPE etc as recommended by the Association for the Advancement of Medical Instrumentation AAMI Standards and Recommended practice Safe Handling and...

Страница 28: ...t immerse the unit overnight 7 Place each component in a critical water e g deionized DI reverse osmosis RO or distilled water bath for a minimum of one minute Do not immerse the unit overnight 8 Usin...

Страница 29: ...econds Minimum Temperature Type of Detergent Water Prewash 2 00 Cold Tap Wash 4 00 Heated Tap 60 C Enzymatic Cleaner per the manufacturer s instructions Neutralize 2 00 Hot Tap Neutral Detergent per t...

Страница 30: ...wo separate components so keep them together with other components and take care not to misplace them Figure 2 Immerse Do Not Immerse 3 With the exception of the hand piece and pneumatic hose immerse...

Страница 31: ...e Do Not Immerse 6 With the exception of the hand piece and pneumatic hose place the components into a fresh tap water bath for a minimum of one minute and then repeat one time using fresh tap water D...

Страница 32: ...th Figure 8 Figure 9 9 Moisten a sterile lint free wipe with critical water and wipe the hand piece and pneumatic hose for a minimum of one minute each Use multiple wipes as necessary CAUTION Do not i...

Страница 33: ...ock nut and lock knob on over the pinion of handpiece to engage at base of handpiece and screw on finger tight counter clockwise till secure NOTE the design on the handpiece should match the lock knob...

Страница 34: ...ned below 9 In a well lit area visually inspect each component in order to ensure all surfaces are clean and dry 10 Using the lock nut wrench replace the locking knob and locking nut back onto the han...

Страница 35: ...not crimp it into the compartment Place the hand piece into the sterilization tray Ensure that the lock nut and lock knob have been previously assembled to the hand piece Use the lock nut wrench to p...

Страница 36: ...elf life is dependent on the sterile barrier employed manner in which product is stored environment and handling conditions A maximum shelf life for sterilized medical devices before use should be def...

Страница 37: ...m fails to spin excision ring blade when activated hand piece when engaged with the disposable excision ring blade and lock ring installed Verify that the excision ring blade and its mating surfaces a...

Страница 38: ...ion ring blade Check the air connection to ensure the pressure setting is 100 psi with the system running allowing the motor to provide continuous operation to the excision ring blade without interrup...

Страница 39: ...cent moon shaped patterns or uneven thickness of grafts Reduce speed and review proper angle and pressure to ensure that the correct technique is being used Skin graft is not uniform or too thick narr...

Страница 40: ...erile mineral oil and exercise the lever in the ON position until it is no longer tacky O Ring on Hand Piece appears worn Normal wear of the O ring is expected The O ring will be replaced during the a...

Страница 41: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 37 of 56 This page intentionally left blank...

Страница 42: ...user guide If the system becomes defective because of misuse or abnormal conditions of operations these repairs shall be charged to the customer All products returned for warranty or non warranty repa...

Страница 43: ...ration warranty and service of the system The Amalgatome SD System must be returned every 12 months for inspection and preventative maintenance The pneumatic hose must be replaced every 12 months or s...

Страница 44: ...to register a complaint Our Customer Service team is available Monday through Friday 8 am to 5 pm ET and is here to assist you in any way Email support exsurco com or call 1 800 243 6049 ext 4 To help...

Страница 45: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 41 of 56 This page intentionally left blank...

Страница 46: ...ome SD System As per internal hospital procedure Weight and Measurement Assembled 2 inch hand piece 3 1 lb Assembled 4 inch Unit 3 7 lb Hand Piece and Pneumatic Hose weight 2 7 lb length 15 feet Compl...

Страница 47: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 43 of 56 This page intentionally left blank...

Страница 48: ...each X100968 Amalgatome SD 2 inch Cutting Head 1 each X100969 Amalgatome SD 4 inch Cutting Head 1 each X100768 Amalgatome SD 2 inch Lock Ring 1 each X100825 Amalgatome SD 4 inch Lock Ring 1 each X1009...

Страница 49: ...each X100966 Amalgatome Lock Ring Removal Tool 1 each X101133 Amalgatome Lock Nut Wrench 1 each X101456 Amalgatome SD User Guide 1 each X101491 Amalgatome SD Intro Letter Part SDSYSTEM04 Description...

Страница 50: ...800 243 6049 ext 4 or email support exsurco com For a price quote and all product related questions please contact your local Exsurco Medical sales representative or contact our Sales department at 1...

Страница 51: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 47 of 56 This page intentionally left blank...

Страница 52: ...r Caution USA federal law restricts this system to sale by or on the order of a physician Use by date This symbol is intended to indicate that the system should not be used after the end of the date s...

Страница 53: ...GATOME SD USER GUIDE PN X101456 Rev C Page 49 of 56 Symbol Definition Caution Consult instructions for use Keep product dry Do not use if package is damaged This side up Sharp object General warning s...

Страница 54: ...f 56 Appendix D Quick Reference Guides Setup Set depth gauge to preferred setting Open package right to left Teeth side down Tighten counterclockwise Sharp excision ring blade do not touch inner ring...

Страница 55: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 51 of 56 Postoperative Procedure Set depth gauge preferred setting Never touch inner ring of excision ring blade...

Страница 56: ...of 56 Manual Cleaning of the Reusable System Clean with a soft bristle brush Lint free wipe Dry Reassemble lock nut onto hand piece Load ASD into Sterilization Tray SOAK WIPE TAP WATER BATH x2 TAP WAT...

Страница 57: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 53 of 56 This page intentionally left blank...

Страница 58: ...Medical Devices in Health Care Settings Validation Methods and Labeling Document Guidance for Industry and Food and Drug Administration Staff Current version of ANSI AAMI ST8 Hospital steam sterilizer...

Страница 59: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 55 of 56 This page intentionally left blank...

Страница 60: ...Amalgatome SD Phone 1 800 243 6049 Website www exsurco com Press the number below for the following departments or email at 1 Sales Clinical Support sales exsurco com 2 Marketing Product Inquiries in...

Страница 61: ...This page intentionally left blank...

Страница 62: ...ad Wakeman OH 44889 Exsurco Medical P O Box 376 Vermillion OH 44089 2017 Exsurco Medical Inc All Rights Reserved Amalgatome and Exsurco are registered trademarks of Exsurco Medical Inc PN X101456 Rev...

Отзывы: