A
MALGATOME
®
SD
U
SER
G
UIDE
PN#: X101456, Rev. C
Page 24 of 56
Manual Cleaning of the Sterilization Tray
1.
Prepare a neutral pH or instrument-safe enzymatic cleaning solution according to the cleaning solution
manufacturer specifications.
Figure 1
2.
Disassemble the tray into its individual
components:
•
Base
•
Lid
•
Internal instrument tray
NOTE:
For additional part number
information, refer to Appendix B: Parts
Information.
3.
Rinse each component under running tap water for a minimum of one minute in order to remove excess soil.
4.
Immerse the base, internal instrument tray, and lid into the cleaning solution and let soak for a minimum of
five minutes. Do not immerse the unit overnight.
5.
While still immersed in the cleaning solution, use a non-metal, soft-bristled brush moistened with cleaning
solution to scrub each component for a minimum of two minutes each and until visibly clean.
6.
Place each component in a tap water bath for a minimum of one minute, and then repeat one time using fresh
tap water. Do not immerse the unit overnight.
7.
Place each component in a critical water (e.g., deionized (DI), reverse osmosis (RO), or distilled water) bath
for a minimum of one minute. Do not immerse the unit overnight.
8.
Using a sterile, lint-free wipe, dry each component. Use multiple wipes as necessary.
9.
In a well-lit area, visually inspect each component in order to ensure all surfaces are clean and dry.
Automated Cleaning of the Sterilization Tray
1.
Prepare a neutral pH or instrument-safe enzymatic cleaning solution according to the cleaning solution
manufacturer specifications.
2.
Disassemble the tray into its individual components:
a.
Base
b.
Lid
c.
Internal instrument tray
NOTE:
For additional part number information, refer to Appendix B: Parts Information.
3.
Rinse each component under running tap water for a minimum of one minute in order to remove excess soil.
4.
Immerse the base, internal instrument tray, and lid into the cleaning solution and let soak for a minimum of
two minutes. Do not immerse the unit overnight.
5.
While still immersed in the cleaning solution, use a non-metal, soft-bristled brush moistened with cleaning
solution to scrub each component for a minimum of one minutes each and until visibly clean.
Содержание Amalgatome SD
Страница 1: ...Amalgatome SD User Guide EXSURCO MEDICAL 10804 GREEN ROAD WAKEMAN OH 44889 PN X101456 REV C...
Страница 2: ...This page intentionally left blank...
Страница 4: ...This page intentionally left blank...
Страница 8: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 4 of 56 This page intentionally left blank...
Страница 14: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 10 of 56 This page intentionally left blank...
Страница 26: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 22 of 56 This page intentionally left blank...
Страница 41: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 37 of 56 This page intentionally left blank...
Страница 45: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 41 of 56 This page intentionally left blank...
Страница 47: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 43 of 56 This page intentionally left blank...
Страница 51: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 47 of 56 This page intentionally left blank...
Страница 57: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 53 of 56 This page intentionally left blank...
Страница 59: ...AMALGATOME SD USER GUIDE PN X101456 Rev C Page 55 of 56 This page intentionally left blank...
Страница 61: ...This page intentionally left blank...