SPECIFICATIONS
Accuracy is specified for a period of one year after calibration and at 18°C to 28°C, with relative humidity < 75%.
Accuracy specifications take the form of: ± ([% of R[number of Least Significant Digits])
Range
Resolution Accuracy
2V
0.001V
± (1.0% + 5)
20V
0.01V
200V
0.1V
250V
1V
± (1.2% + 5)
Range
Resolution Accuracy
200mV
0.1mV
± (0.8% + 5)
2V
0.01V
20V
0.01V
200V
0.1V
250V
1V
±(1% + 5)
Input Impedance:
10MΩ
Frequency Range:
40Hz ~ 400Hz
Overload Protection:
250V DC/AC rms
Response:
Average, calibrated in rms of sine wave
Max. Input Voltage:
250V AC rms
Range
Resolution
Accuracy
200µA
0.1µA
± (1.2%+5)
2000µA
1µA
20mA
0.01mA
200mA
0.1mA
2A
0.001A
± {2.0%+10}
10A
0.01A
Overload Protection:
µA
and mA ranges: F 250mA/250V fuse
A ranges: F 10A/250V fuse
Max. Input Current:
«INPUT» jack : 200mA
«10A» jack: 10A
( For measurements > 2A: duration < 15 seconds,
interval > 15 minutes )
Voltage Drop:
200µA, 20mA and 2A ranges: 20mV
2000µA, 200mA and 10A ranges: 200mV
Range
Resolution
Accuracy
200µA
0.1µA
± (1.5%+5)
2000µA
1µA
20mA
0.01mA
200mA
0.1mA
2A
0.001A
± {3.0%+10}
10A
0.01A
Overload Protection:
µA
and mA ranges: F 250mA/250V fuse
A ranges: F 10A/250V fuse
Max. Input Current:
«INPUT» jack : 200mA
«10A» jack: 10A
( For measurements > 2A: duration < 15 seconds,
interval > 15 minutes )
Voltage Drop:
200µA, 20mA and 2A ranges: 20mV
2000µA, 200mA and 10A ranges: 200mV
Frequency Range:
40Hz ~400Hz
Response: Average, calibrated in rms of sine wave
Range
Resolution
Accuracy
meter
200A
0.1mV/0.1A
± (1.2%+5)
meter
1000A
1mV/1A
± (1.2%+5)
Max. Input Voltage:
200mV
DC Current (with clamp)
AC Current
DC Current
DC Voltage
Input Impedance:
10 MΩ
Overload Protection:
250V DC/AC rms
Max. Input Voltage:
250V DC
AC Voltage