Manual UVC Air Sanitizer V1000E
13.11.2020
Page
18
of
33
6.5.2 Operating via network
The communication is set up via TCP/IP connection. Each device has a static IP address. The
default IP Address is:
192.168.000.010
Communication port is HTTP port 80.
The command and the reply from the device are formatted as JSON object.
Command
Param
Answer
Value
Note
state
none
statePower [value]
stateTubes[value]
stateFans [value]
fanspeedPct[value]
tubesHours[value]
alarmTubes [value]
alarmFans [value]
fanspeedRpm[value]
currentTubes[value]
serial[value]
fw[value]
“1” or “0”
“1” or “0”
“1” or “0”
“50” to “100”
0.0 to …
“1” or “0”
“1” or “0”
0 to …
0 to …
SerNr
0.0 to …
Power on or off
UVC tubes on or off
Fans on or off
Fan speed in percentage
UVC tubes operating time
UVC tubes alarm
Fan alarm
Fan speed in rpm
Tube current as relative value
Serial Number as string
Firmware
setFanSpeed
[param]
[50] to [100]
fanspeedPct [value]
“50” to “100” Setting the fan speed in
percentage
setPower
[param]
[1] or [0]
statePower [value]
“1” or “0”
Setting power on or off
setIP [param] [new IP
Address]
Status message
[value]
“Trying to
change IP
address”
IP address formatted as:
AAA.BBB.CCC.DDD
Example: 192.168.000.010
Table 2 Description of API
tubesHours: Operating hours are counted in 15-minute steps.
Decimal places are part of an hour, not minutes.
Example: An answer from the device of 5.75 is equal to an operation time of 5h45m.
Note: The network setting can be reset to default values by pressing the
multifunctional switch for approx. 20 seconds.