6 Operation
Instructions for use
•
MERGITE!
51/84
Note
>
The wash buffer used in the container should not be used for
more than
1 week
. We recommend producing new wash buffer
at the start of the working week.
>
Ensure that as little air as possible is introduced into the wash
buffer when wash buffer is applied. Increased bubble formation
can have negative effects on the washing result.
Chapter 9.2 Liquid system, p. 74
1) Place a paper towel at the ready to mop up any dripping liquid
residues.
2) Disconnect the container.
3) Fill the empty container with wash buffer.
4) Reconnect the filled container.
5) If you were prompted by the system to fill the wash buffer
container, finish the process with
OK
.
6)
Rinse with the new wash buffer.
Rinsing with wash buffer, p. 51
You filled the wash buffer container and rinsed the system with wash
buffer.
6.5 Rinsing with the wash buffer
Rinsing with wash buffer
The device must be rinsed with wash buffer when
the device is switched on.
the device was switched on and has been inactive for a long time.
the wash buffer container was refilled.
when air bubbles or foam build-up is observed on the wash
columns.
1) When you were prompted by the system to rinse with wash
buffer, continue with
2) Open the
RINSE
view.
START
>
MAINTENANCE
>
RINSE
or as alternative for thorough rinsing
START
>
MAINTENANCE
>
FULL
RINSE