8 Maintenance
Instructions for use
•
MERGITE!
67/84
8.3 Maintenance and cleaning tasks
The maintenance and cleaning tasks are detailed in the following.
Filling the system with distilled water
Turn off the device if it is anticipated that the device will not be used
for an extended time (e.g. overnight). In this case the liquid system is
filled with distilled water in order to prevent salt deposits. This
significantly increases the service life of the valves and pumps.
Switching the device off, p. 55
8.3.1 Cleaning of the liquid system
Regular cleaning of the liquid system is necessary to remove any
biological deposits in the hoses, pumps, etc.
Cleaning the liquid system
The following instructions describe the cleaning of the liquid system
step-by-step.
ATTENTION
DAMAGE FROM CLEANING AGENTS
The liquid system can be damaged by overly long exposure to the
cleaning agent, the wrong concentration or the use of an incorrect
cleaning agent.
>
Always comply with the exposure times.
>
Use only SETUP CLEAN (1:5).
You will require two containers with SETUP CLEAN (1:5).
Chapter 8.3.4 Container preparation with SETUP CLEAN, p. 72
1) If you are prompted to perform monthly maintenance when you
start the device, press
START
and then continue with step
2) Press
START
>
MAINTENANCE
, in order to open the
MAINTENANCE
view.
3) Press
MONTHLY
MAINTENANCE
.
You will be prompted to connect the containers with
SETUP CLEAN (1:5) to connectors In 1 and In 2.