![EUROIMMUN Medizinische Labordiagnostika AG MERGITE! Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/euroimmun-medizinische-labordiagnostika-ag/mergite/mergite_instructions-for-use-manual_2437749078.webp)
10 Technical data
78/84
Instructions for use
•
MERGITE!
10.3 Electrical and operating data
External AC adapter (mains adapter)
Only use the provided mains adapter and mains cable.
AC/DC mains adapter ............... autoselection 100-240 V AC / 47 - 63 Hz / DC 24 V, 100 W
Device
Supply voltage .......................................................................................................... 24 V DC
Power consumption ..................................................................................................... 100 W
Protection class of device .............................................. III - safety extra-low voltage (SELV)
Protection class according to EN 60529 ........................................................................ IP 20
Noise/sound pressure level ..................................................................................
≤ 65 dB(A)
Interference emission .......................................... according to EN 55011, Group 1, Class
A
FCC Part15-B, Class A
Electromagnetic compatibility (EMC) .................................................................. ANSI C63.4
Interference immunity ........................................................................ according to EN 61326
Electrical safety ................................................ according to EN 61010-1 / EN 61010-2-101
EN 61326-1 / EN 61326-2-6
Overvoltage/installation category .................................................................. II (EN 61010-1)
Contamination level ....................................................................................... 2 (EN 61010-1)
Maximum pump pressure (connector In 1, In 2, Out) .................................................... 1 bar
The product satisfies the requirements regarding interference emission and immunity in
accordance with the EN 61326 series of standards and fulfils the limit values in accordance
with DIN EN 55011, Group 1, Class A (operation in an industrial environment). The product
was developed and tested in accordance with CISPR 11 Class A. It is not intended to be used
in residential areas and cannot ensure adequate protection of radio reception in such
environments.
Device connections
DC In .............................................................................................................. supply voltage
Service (USB-B) ........................................................... service interface (connector for PC)
Assay (USB-A) ............................................................................. connector for data carrier
In 1, In 2 ..................................................... quick release plug (distilled water, wash buffer)
Out .............................................................. self-closing quick-release socket (waste liquid)
M8 socket (4-pin) to In 1, In 2 and Out ..................................... fill level detection connector