![EUROIMMUN Medizinische Labordiagnostika AG MERGITE! Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/euroimmun-medizinische-labordiagnostika-ag/mergite/mergite_instructions-for-use-manual_2437749042.webp)
6 Operation
42/84
Instructions for use
•
MERGITE!
11) Press
START
.
Note
The device automatically checks the fill level of the container. If the
fill levels are not sufficient for the washing process, you are
automatically prompted to change or refill the corresponding
container.
Chapter 6.4 Filling or draining the containers, p. 49
The washing of the slides begins.
12) At the beginning of the washing process, make sure that all
substrate fields have contact with the washing liquid.
If necessary, you can establish the contact by tapping on the
corresponding field with your finger.
The duration of the wash process depends on the assay selected.
Note
If you remove the slide carrier from the wash basin during the
washing process, the washing process is automatically interrupted.
Afterwards, an aspiration process is carried out to remove liquid
residue from the slides.
Fig. 19 Slide washing with MERGITE!