13.12.3
Transmission of limits
If you select or edit a new parameter set in which limits are entered and if you have selected the DP-1 HS as printer (see
previous section), the system will prompt you to transmit the limits to the printer. When not, reselect the selected parameter
set.
If you answer with "yes", then follow the instructions on the GMV2-K screen:
The old limits are printed out when you press the CL followed by the TOL.LIMIT key. Wait until the printout is finished,
then quickly confirm with "yes" on the GMV2-K. Now press the data key on the printer. The "NO DATA" error message
may appear. Please ignore this message!
Then acknowledge with" yes”, press the TOL.LIMIT key again and confirm with "yes" on GMV2-K. The transmission is
thus terminated; the new limits will be printed out.
Notes:
•
Wait until the printout is finished before hitting the next key on the GMV2-K!
•
Once the transmission is started it cannot be aborted!
13.12.4
Printing out the measured values
The following printout options are also available with the DP-1 HS setting:
-
Single-value printout
-
Printout over erasure time (setting: "RS-232")
-
Printout one or more data records from a single parameter set (you may need to press the CL key first)
Notes:
•
The printer can only process one numeric range. Should the position of the decimal point or the sign change when a
new parameter set is selected, the CL key must be pressed before the first printout.
•
If torque and angle of rotation are measured, only the torque can be printed out.
•
The
parameter sets
can still be output to a standard printer or to the PC with the
appropriate cable
! The baud rate
must be previously selected with the "Standard" setting.
•
Most statistics functions require limits. They should be transmitted first.
•
Torque values are printed out without units.
•
No
angle values are printed out!
14 Error messages
14.1 General
The GMV2-K has been designed so that operator errors are almost impossible. Invalid settings are not displayed when
selected and inhibited keys have no inscription.
A plausibility check is performed during parameter set programming so that, for example, the minimum value cannot be set
greater than the peak. The system also ensures that no over-range is possible. An incorrect entry is indicated with an error
message and, if possible, the value is corrected. Invalid operating states are also displayed.
An error message is output at the left-hand side of the LCD display. A beep is also sounded.
There are three different types of messages:
•
Notes: For minor faults. This error message is automatically cleared after one second.
•
Warnings: One or more functions cannot be started. All GMV2-K functions are stopped until this message is cleared
with the return key.
•
Major faults: The GMV2-K is not functioning properly. The unit must be switched off and the fault cleared