ETH-messtechnik gmbh
D-74415 Gschwend P.O. 16
Nr.: 162e
Operating Instructions GMV2-K
revised: 09.05.2019
Rev.: 0
Subject to change without prior notice
32
Tip:
You can overwrite individual measured values before transmission by selecting another cell. The selected cell must be
in
the first row
of the cell that you want to overwrite. Now you can transfer single measured values, a single data record, or
all data records to the GMV2-K.
In “Menu – Output – RS232 – Store” you can select which data you wish to transmit. You can then see the measured values
being scanned into the table on the PC.
When transmission is complete, you need to acknowledge this by clicking on “Finished?”
If you transmit “all” measured values (all measured values or those from one parameter set), the transmission is
automatically terminated. The button then changes back to “Scan” and the measured-value table is processed:
•
empty rows are hidden, used rows are shown
•
the measured values are divided according to parameter set und data record numbers by means of dividing bars
•
NOK measured values (outside the limits) are stored in red
•
Print area und page are automatically set up (!)
13.4.4 Measured-value table
Some of the unused rows may be hidden in the measured-value table. This happens automatically with rows “B” to “L”..
Leave the row “M” empty, as it separates the subsequent rows of comments. You can use the rows of comments as you
will: hide or show them and use them for notes for reading in measured values. They are not altered by the program
, and
typically do not appear in the printout!
The print area is automatically set for all measured values. Should you wish to print out some of the data only, change the
print area beforehand.
Tip:
•
If you wish to print comments as well, you can enter any character in the first line in the “M” row and then set the row
width to zero. A comments row is only hidden when it is completely empty.
13.4.5 Statistical analysis
In the “Statistics” page, you can select
–
according to the parameter set and the data record
–
which measured values
should be used in the calculation and shown in the diagram.
You can adjust the head of the page, up to the selection fields, as you wish. Do not alter the section containing the statistics!
Although the print area is pre-defined on this page, it can still be altered.
Note:
•
When both limit values are negative,
counter-clockwise rotation
is assumed.
A measured value that is more negative than the limit value, will then be interpreted as being
too large!
Similarly, a
more positive value is too small!
13.4.6 Measured-value diagram
The data range for the first row of measured values (and for the limits) is automatically set in the dropdown selection
fields. Scaling is also set automatically. You can, however, enter values for (fixed) manual scaling.
You may arrange the diagram (colors, line thickness, ...) as you wish. The best way to do this is to right-click on the desired
element (context menu).
Note:
•
By pointing the mouse exactly over a vertex of the curve, the numerical figures will be shown.
•
You may insert more self-defined diagram (and table) sheets.
13.4.7 Torque wrench document
This page is intended as a starting point for your own test logs. It accesses the measured-value table via formulas. This is
why you can alter this page as you wish.
Содержание GMV2-K