ETH-messtechnik gmbh
D-74415 Gschwend P.O. 16
Nr.: 162e
Operating Instructions GMV2-K
revised: 09.05.2019
Rev.: 0
Subject to change without prior notice
23
9.6 Selection
Menu path:
Menu-> Parameter -> ParSet-> choose
A parameter set becomes the current set when selected and confirmed with "OK". The measurements will be made with
these settings. The number of the selected set is displayed on the top left of the screen to the left. The "return" key aborts
the selection!
Notes:
•
The parameters are copied internally to set number "0" with this function. Direct entries are thus overwritten.
•
When working with switchable transducers with numerous measuring ranges and measuring-range detection per chip,
the desired parameter set must be programmed and e.g. selected in GMV2-K before the measuring range can be
changed over. Then, the error message is removed and the measuring level is reached by pressing the Return key.
•
When you are working with transducers with two measuring ranges the transducer is switched automatically to the
desired measuring range by entering the required measuring channel in the parameter set.
•
If the DP-1 HS (optional) is selected as printer and the parameter set contains torque limits, the system prompts you to
transmit the limits to the printer.
9.7
Create new
Menu path:
Menu-> Parameter -> ParSet-> new
To quickly set all necessary parameters for the analyzing unit, a parameter set should be created for this purpose. GMV2-K
can store 49 different sets. The sets are not lost when the unit is switched off.
A new parameter set is automatically assigned the number of the first free memory area. Gaps arising when individual sets
are erased can thus be re-assigned. You may now start programming ("program" menu) or use an existing parameter set as
a template ("Copy" menu). A set, including set number "0", can be selected for copying with "Param - " and "Param +".
After confirmation with "OK" you should start programming. All settings are displayed and the values are checked. A
parameter set is programmed in the same way as with editing. A more detailed description is given in the following section.
Notes:
•
When a parameter set is first created, it is automatically selected and becomes the current set.
•
If the DP-1 HS (optional) is selected as printer and the parameter set contains torque limits, the system prompts you to
transmit the limits to the printer. The printer has to be connected in this case.
9.8
Editing
Menu path:
Menu-> Parameter -> ParSet-> change-> edit
Settings in existing parameter sets may be altered in this menu. However, if measured values have already been stored with
this set it is protected against changes and an error message will be output. Before a stored parameter set can be used,
measured values stored in this parameter set in the device have to be deleted.
To ensure that all parameters are entered, the user is forced to enter the parameters in succession and to confirm with "OK".
This entry loop cannot be aborted prematurely! However, you can always confirm with "OK" and thus quickly reach the
end of the loop. In which case you must also confirm the last prompt "All entries correct? "" with with "yes". Otherwise the
loop will run through again.
Only the necessary parameters are prompted depending on the selected measuring mode and can be set with the Entry
editor (Sec. 8). The order has been arranged so that a plausibility check of interdependent values is made. If entries are
incorrect, an error message is displayed after confirmation, an acoustic alarm is output and the value is automatically
corrected so that it is within the permissible range!
Some queries can be skipped by having fixed values in the defaults and by hiding functions (see secs 3.10.1 and 3.10.2,
page 11).