26
ENGLISH
product.
product).
The window width sets the distance from the setting up point where the 4-
20mA mode starts: 4mA = 0 metering pump pulse, 20mA = max frequency of
the metering pump. The window width depends on many variables: distance
of the injection point, reaction time, chemical product concentration, etc.
With the overdose time alarm one can select a period during which the setpoint must be reached. If the setpoint
is not reached during this period of time, the instrument blocks output operations, including those in mA
(metering pumps), the alarm is displayed as ON and triggers a signalling instrument connected to the alarm
relay.
Maximum Metering is a safety function that ensures metering is completed in
the selected time. The relays of the metering pumps open accordingly. This
function makes it possible to eliminate time limits (continuous metering
according to the selected setpoints) or to select the minutes (up to 999) and
the hours, for instance metering for 999 minutes in 24 hours.
Press E S C to go back to M E N U S E L E C T I O N or press E S C E S C to go back to M E A S U R E M E N T D I S P L A Y
A
CALIBRATION > EXPERT MENU
In the EXPERT MENU mode, calibration does not change. Therefore, refer to the CALIBRATION menu – BASIC MENU, illustrated
above.
START/STOP TIMES > EXPERT MENU
In the EXPERT MENU mode, the START/STOP TIMES menu does not change. Therefore, refer to the START/STOP TIMES menu
– BASIC MENU, illustrated above.
AUX OUTPUT > EXPERT MENU
◄
►
The auxiliary output controls various functions connected to any type of remote On-Off device
controlled by a timer in real time. The AUX output may control a device or appliance thanks to very
accurate programming of minutes/hours/days/weeks. ADVANTAGES: this function makes this
instrument a very versatile control unit not only to measure chemical physical parameters but also for
other functions connected to the system where it is installed.
Press
O
O K / M E N U
U
to confirm the selection and to go on to the next sub-menu.
In the highlighted steps
select the desired parameters; confirm the selection with
O
O K
K
; move on to the next steps.
Below are the programming steps to be selected:
The settings of the program entail many steps, which may be selected to have absolute operating precision
(up to 99 programs for each AUX output).
SETTING 4-20mA (pH)
Metering
Alkaline
EXPERT MENU
SETTING 4-20mA (RX)
Metering
Reverse
EXPERT MENU
SETTING 4-20mA (pH)
Window width
1.50 pH
EXPERT MENU
SETTING 4-20mA (RX)
Window width
150 mV
EXPERT MENU
SETTING 4-20mA (pH)
Overdose
00.00 h:m
EXPERT MENU
SETTING 4-20mA (pH)
Max. Metering
000 (no limit)
EXPERT MENU
SETTING 4-20mA (RX)
Max. Metering
000 (no limit)
EXPERT MENU
Menu Selection
Aux Output #1
AUX 1
Program
01
Active time (m:s)
01 : 00
Start time (h:s)
10 : 30
Enabl. days
Mo:
N
Tue:
N
Wed:
N
Thurs
:
N
Fri
:
N
Sat
:
N
Sun:
N
Wk.
1
:
Y
2
:
Y
3
:
Y
4
:
N