![ETATRON D.S. eSelect M 1 pH (Rx) Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/etatron-d-s/eselect-m-1-ph-rx/eselect-m-1-ph-rx_operatinginstructions-and-maintenance_2431288047.webp)
20
ENGLISH
Rx (REDOX) CALIBRATION > BASIC MENU
* When the Calibration starts, the display shows random values
Short-circuit the BNC by using a copper wire, for calibration only, to put in contact
the internal PIN with the external part of the connector.
Press ◄► and ensure the display shows
0
0 mV
V
Press
O
O K / M E N U
U
to confirm the selection
Dip the RX electrode in the buffer solution for “Gain” calibration, wait 2/3 minutes.
Press ◄► and ensure the display shows the mV value relating to the chosen buffer solution.
Press
O
O K / M E N U
U
to confirm the selection
Press E S C to go back to the M E N U S E L E C T I O N
U
Press E S C E S C to go back to the M E A S U R E M E N T D I S P L A Y
A
The mV value at the bottom shows the electrode signal and consequently its efficiency (according to the manufacturer’s specifications).
Refer to the electrode’s instructions and ensure the mV value displayed matches that of the buffer solution with a tolerance of
20mV.
START/STOP TIMES > BASIC MENU
The Timer in real-time programs the days and exact time of instrument operations.
Program the date and time by selecting the backlit steps.
Press ESC to go back to MENU SELECTION or press ESC ESC to go back to the MEASUREMENT DISPLAY
A
SETTINGS > BASIC MENU
BASIC programming: simplified mode for end user or non-professional
operators
EXPERT programming: complete programming that includes functions for
more accurate control and results (for the installer).
Program the date and time by selecting the backlit steps.
CALIBRAT.
MEAS.1 (RX)
Cal. Zero
0 mV
27.5 mV
CALIBRAT.
MEAS.1 (RX)
Cal. Gain
450 mV
420.1 mV
Menu Selection
Start/Stop times
Start/Stop times
Monday
00:00
◄
–
►
24:00
EXPERT MENU
Start/Stop times
Tuesday
00:00
◄
–
►
24:00
EXPERT MENU
Menu Selection
Settings
Settings
Menu Mode
BASIC
MENU BASE
Settings
Menu Mode
EXPERT
MENU BASE
SETTINGS
Date
2017 May 14
BASIC MENU
SETTINGS
Time
12:02
BASIC MENU