16
ENGLISH
INITIAL DISPLAY
NOTE FOR THE PROGRAMMER: Read the manual before starting programming or always have it at hand to be sure you are making the
correct selections.
IMPORTANT: if no keys are pressed for 60 seconds, the instrument will show the current measurement.
To go forward quickly, press and hold one of the ◄► buttons
The software version is shown when the instrument is on the lower part of the display.
The software is subject to revisions without notice.
The instrument is prepared for measuring and is then ready to operate.
At this stage some status messages might be displayed such as:
S1 S2 L1 DATE/TIME AUX OVERDOSE ALARM MAX DOSE FLOW BASIC MENU EXPERT MENU
which might be active due to the current measurements, just go Forward with programming.
The INITIAL DISPLAY shows the measurements according to the M easurement Type selected in the
paragraph.
SELECT MEASUREMENT TYPE PH or RX”
. If the instrument has already been programmed,
the display shows the programs selected previously.
NOTE: if the RX M easurement Type is selected, all values are automatically in mV.
The MAIN DISPLAY shows the measurements of the parameter, the status of the functions and the alarm
indications.
When the instrument is switched on for the first time, a list of functions of all keys is displayed. This screen
is no longer displayed during subsequent start-up operations.
READ THE MESSAGE CAREFULLY, THEN PRESS O K
K
TO START.
USE OF THE KEYS
OK: access to the menu, start/stop, selection and editing
selects the step of the menu or increases / decreases the value
ESC: goes back to the previous menu or does not save the change
* Press OK to start*
To go forward quickly, press and hold one of the ◄► buttons
IN ENGLISH
Select the language of th e PROGRAMMING MENU
To select the language of the programming menu.
After selecting the Language (ITALIAN, ENGLISH), the programming menu
adapts accordingly.
Select MEASUREMENT TYPE pH or Rx
The instrument of the eSelect M PH(RX) series makes it possible to select the type of measurement to be used, i.e. PH or RX
(Redox). The measurement parameter may be changed from PH to RX and the display and the measurement resolutions change accordingly.
◄
►
◄
►
IMPORTANT: after selecting the Measurement Type PH or RX, ENSURE you connect the appropriate electrode.
ETATRON
Model eSelec M1
R ev . X. X
CONTROLLER SETTINGS
Language
English
BASIC MENU
SETTINGS
Lingua
Italiano
MENU BASE
Measurement Type
pH
Measurement Type
Rx