With the Overdose alarm one can select a period during which the setpoint must be reached. If the setpoint is
not reached during this period of time, the instrument blocks output operations, including those in mA (metering
pumps), the alarm is displayed as ON and triggers a signalling instrument connected to the alarm relay.
Maximum Metering is an extra safety function that ensures metering is
completed in the selected time. This function makes it possible to eliminate
time limits (continuous metering according to the selected setpoints) or to
select the minutes (up to 999) and the hours, for instance metering for 999
minutes in 24 hours (see MAXIMUM METERING TIME menu).
The start-up delay stops the output relays when the unit is switched on, thus allowing the sensor to polarise
assuring correct measurements (programmable).
ONLY WHEN USING METERING PUMPS
The priming function blocks the setpoint value to allow the metering pump to
remain primed, the outputs are not active.
Press E S C to go back to M E N U S E L E C T I O N or press E S C E S C to go back to M E A S U R E M E N T D I S P L A Y
A
SETPOINT 2 PH AND SETPOINT 2 RX > EXPERT MENU
Refer to the SETPOINT 1 menu, as the SETPOINT 2 menu has the same
settings as the SETPOINT 1 menu, with the exception of the MIN/MAX Alarm
function, indeed only found in the SETPOINT 1.
ANALOGUE 4-20MA OUTPUTS FUNCTION SELECTION > EXPERT MENU
◄
►◄
►
Press O K / M E N U to go into mA mode and choose between two operative functions:
◄
►◄
►
-
mA DEVICE: external devices suited to operating with mA signal such as data loggers or data recorders or other mA equipment
-
SETPOINT DISPENSING: through the control of metering pumps suited to controlling an mA external digital signal.
4-20mA1 / 4-20mA2 pH or Rx OUTPUTS > REMOTE DEVICES FUNCTION
MENU SETPOINT 1 pH
/
RX
Overdose
00.00 h:m
EXPERT MENU
MENU SETPOINT 1 pH
/
RX
Max. Metering
000 (no limit)
EXPERT MENU
MENU SETPOINT 1 pH
/
RX
Max. Metering
000 m / 00 h
EXPERT MENU
MENU SETPOINT 1 pH
/
RX
Start-up Delay
00.00 h:m
EXPERT MENU
MENU SETPOINT 1 pH
/
RX
Priming
OFF
EXPERT MENU
MENU SETPOINT 1 pH
/
RX
Priming
ON
EXPERT MENU
Menu Selection
Setpoint 2 (pH)
Menu Selection
Setpoint 2 (RX)
Menu Selection
4-20mA1
Menu Selection
4-20mA2
SETTING 4-20mA
mA output mode
mA Device
EXPERT MENU
SETTING 4-20mA
mA output mode
Setpoint
metering
EXPERT MENU