P R E P A R I N G T H E S Y S T E M
MyLab
- G E T T I N G S T A R T E D
4 - 4
Fig. 4-2: B) Medical Use Room
Intended as area B, area A excluded.
Fig. 4-3: C) Non-medical use room
A room not designed for medical treatment, for example, an office or a
storage room.
Possible configurations:
MyLab
+ auxiliary device complying to IEC 60601 in area A
- No additional safety requirements.
MyLab
+ auxiliary device not complying to IEC 60601
(complying to IEC XXX
1
) in area A
- Auxiliary device must be powered through a safety
insulation transformer complying to IEC 60601.
1.
IEC XXX stands for standards such as: IEC 60601 for medical devices, IEC 60950 for
information technology equipment etc.