A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y
MyLab
- G E T T I N G S T A R T E D
2 - 6
Electromagnetic Immunity for All Medical Equipment
MyLab
is intended for use in the electromagnetic environment specified below. The customer or
the user of
MyLab
should assure that it is used in such an environment.
Immunity Test
IEC60601
Test Level
Compliance
Level
Electromagnetic Environment
and Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
kV on contact
kV in air
kV on contact
kV in air
The floor should be in wood,
concrete or ceramic tiles. If floors
are covered with synthetic
material, the relative humidity
should be at least at 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines
2 kV for power
supply lines
1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
mode
1 kV differential
mode
2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips
IEC 61000-4-11
0%
U
T
; 0,5 cycles
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at
0°
0%
U
T
; 0,5 cycles
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at
0°
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of
MyLab
requires continued
operation during power mains
interruptions, it is recommended
that
MyLab
is powered from an
uninterruptible power supply or a
battery.
Power interruptions
IEC 61000-4-11
0%
U
T
; 250/300
cycles
0%
U
T
; 250/300
cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of
MyLab
requires continued
operation during power mains
interruptions, it is recommended
that
MyLab
is powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Содержание MyLab XPro75
Страница 1: ...Rev 03 June 2021 MyLabX75 Family GETTING STARTED 350079900...
Страница 2: ......
Страница 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity...
Страница 6: ...MyLab G E T T I N G S T A R T E D vi...
Страница 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED...
Страница 8: ...MyLab G E T T I N G S T A R T E D viii...
Страница 9: ...MyLab G E T T I N G S T A R T E D ix GETTING STARTED...
Страница 10: ...MyLab G E T T I N G S T A R T E D x...
Страница 11: ...MyLab G E T T I N G S T A R T E D xi GETTING STARTED...
Страница 12: ...MyLab G E T T I N G S T A R T E D xii...
Страница 30: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Страница 67: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED...