8.5
Checking and Setting the Supply Voltage
On delivery, the unit is set for the local supply voltage (110/115 ~ or 230/240 V ~). The current voltage
setting is recorded on the mains module. The power unit covering lid on the bottom of the unit must be
opened to convert the voltage. The voltage can then be changed on the power unit on-board with a special
voltage selector switch.
Voltage conversion should only be performed by authorised and trained personnel.
8.6
Changing a Mains Fuse
The fuse switch is located in the centre of the mains module. The lid can be pried out of its lock-in position
with the help of a small screwdriver. It can then be pulled out of the fuse well. Two fuses are located in the
fuse switch. After a continuity check, change the fuses if necessary. Return the fuse switch to the well and
press it into the lock-in position.
Replace fuses only with other fuses of the same type with the same electric data: 2 x 1.25 AT for
230 V, or 2 x 2.5 AT for 110 V.
8.7
Disposal Information
Devices that are no longer usable can be returned to ergosana for disposal. Alternatively, the device can be
taken to an approved disposal location.
The control console of the device contains a buffer battery which must be disposed of separately.