Sana Sedeo 150 HK and 250 HK
Part No.: 2. 520038 Rev.: d
Page 18 of 26
7.4
Training program (option)
The use of the training program requires a pulse signal receiver (Polar system), which is integrated in the
control panel. The receiver can be ordered together with the device or added later on.
The patient wears a transmitter belt, which is positioned on the skin below the chest. The effective radius of
the signals transmitted by the belt and received in the control panel is approx. 70 cm. Please note that if the
skin is dry, contact problems may occur between the belt and the skin at the beginning of the training.
Therefore, if the pulse transmission is unsteady or disturbed, moisten the contact surfaces of the transmitter
belt with contact spray or water.
7.4.1
Training with constant heart rate (Pulse-Steady-State)
A bicycle ergometer training with constant heart rate (pulse-steady-state method) in the individual training
range is a highly efficient and risk-free training method for the cardiopulmonary system. Ask your physician
which is the suitable training heart rate for you. He or she will determine it for you using an exercise test.
7.4.2
Configuring a training program on the ergometer
To define a training program, select the menu option
Define program. Then select the menu option Training using the arrow keys.
The below settings panel is displayed.
To set or change the parameters in order, press modify.
The first setting defines
the Initial load. e.g. 50 watts.
The setting Duration A1 controls the du-
ration of warming-up phase 1,
e.g. 1 min.
The setting Increase in load controls by
how many watts per minute the load is
increased.
This is warming-up phase 2. Using the
next setting, Duration A2, this phase
can be limited in time in special cases.
By default, the warming-up phase 2 is
finished when the target HR is reached.
When the target HR is reached, the load
applied is automatically reduced by
10%.
This prevents an "excessive increase in
heart rate" during the training phase
(TP) which now begins.
The Target HR is set by the next pa-
rameter. The setting Duration of training controls the duration of the training phase (TP). The setting
Cool down controls the period of time in which the load is reduced to 0 watts after termination of the
training phase. The following settings Ask for Po – no/yes, Ask for HR – no/yes and Ask for weight –
no/yes can be viewed when a new training is started and edited. Editing these parameters can be re-
quired when different persons want to use the training program.