6
•
Maintenance and servicing
36
/ 42
801
16-
221
10.
15
Test setup
Fig. 6-2
•
The test specimen is connected to the power supply via the power
cord.
•
The Bipolar receptacle of the test specimen is connected to the VIO
Testbox via the bipolar cable with the adapter cable.
Test procedure
1. Activate AUTO START if necessary. To do so:
– Call up service level (see page 19)
– Use the Plus/Minus button to select and the yellow selection but-
ton to activate <Setup 27>.
– Use the top right selection button to set <AUTO ON>.
– Save the setting. Press and hold the top left selection button until
a check mark in the adjacent display indicates the end of the pro-
cess.
– Exit Service mode. Switch the test specimen off.
2. Switch the test specimen on and set to:
BIPOLAR SOFT, 50 Watt, AUTO START
3. Press and hold the T1 button on the VIO test box. The test specimen
must start the activation.
4. Press button T2. The test specimen must terminate activation.
Testbox 20100-102
2
X
X
X
X
1
BP1
BP2
T1
T2
+
+