6
•
Maintenance and servicing
34
/ 42
801
16-
221
10.
15
Activation tones
1. Press CUT pedal on the dual-pedal footswitch at least twice. When
pressing the pedal there must be an acoustic signal every time
(=acknowledgement tone).
2. Press COAG pedal on the dual-pedal footswitch at least twice. When
pressing the pedal there must be an acoustic signal every time
(=acknowledgement tone).
Footswitch activation
Testing and measuring equipment
Test setup
•
The test specimen is connected to the power supply via the power
cord.
•
The dual-pedal footswitch is connected.
Test procedure
Call up
test program
1. Call up service level (see page 19).
2. Use the Plus/Minus button to select and the yellow selection button
to activate the test program <Setup 26>.
Dual-pedal footswitch
activation CUT / COAG
1. Press the CUT pedal on the two pedal footswitch. The "X" character
must appear in the left segment in the CUT display
1
during activa-
tion.
2. Press the COAG pedal on the two pedal footswitch. The "X" character
must appear in the right segment in the CUT display during activa-
tion.
3. Remain in the Test program.
Erbe Art. No.
Description
20189-107
Two pedal footswitch VIO C
1. Legend: "–" not yet activated, "X" successful activation, "0" = activation finished
Содержание VIO 100 C
Страница 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Страница 2: ......
Страница 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Страница 6: ...Table of Contents 6 42 80116 221 10 15...
Страница 20: ...4 Service mode 20 42 80116 221 10 15...
Страница 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...