Remedying malfunctions
801
16-
221
10.
15
CHAPTER 5
Remedying malfunctions
Safety information
Error messages
ATTENTION!
Adjustments, technical tests, modifications, maintenance and repair
work may only be performed by Erbe or persons trained by Erbe. If the
work is not performed by trained persons, Erbe accepts no liability and
warranty rights become void.
Error
code
Description
Additional informa-
tion
Action
1
Reaching the activation period limit.
If the activation period limit has
been set too short, service
program 22 can be used to
extend it.
2
NE monitoring returns an illegal value
at or during monopolar activation.
Check whether the NE is applied
correctly and connected to the
unit.
Check whether the NE cable is
connected correctly and in good
condition.
If the error still occurs, notify
Erbe Service.
3
Activating the unit without prior
acknowledgement of the front panel
setting.
Press any button to acknowl-
edge the settings of the front
panel and then reactivate the
unit.
4
Activation of CUT with no defined
power setting.
Select CUT power setting
> 0 watts before activating the
unit.
5
Activation of COAG with no defined
power setting.
Select COAG power setting
> 0 watts before activating the
unit.
Содержание VIO 100 C
Страница 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Страница 2: ......
Страница 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Страница 6: ...Table of Contents 6 42 80116 221 10 15...
Страница 20: ...4 Service mode 20 42 80116 221 10 15...
Страница 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...