HF power output CUT
Testing and measuring equipment
Test setup
Fig. 6-3
•
The test specimen is connected to the power supply via the power
cord.
•
The levels of power are determined with the HF power meter. The
measuring cables are plugged into the HF power meter direct.
AUTO CUT
DRY CUT
Test procedure for VIO 50 C
1. Set the test specimen to:
AUTO CUT, 50 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via CUT button on the electrode handle.
Erbe Art. No.
Description
20192-127
Patient cable AE
or
Patient cable AE, international
20192-110
–
Laboratory measuring cable
20190-045
Electrode handle ICC/ACC
20194-070
Patient cable NE
or
Patient cable NE, international
20194-075
20100-033
Adapter cable NE
–
HF power meter
20189-107
Two pedal footswitch VIO C
AE
NE
HF power meter
+
+