4
•
Service mode
16
/ 42
801
16-
221
10.
15
Service overview
ATTENTION!
In this table, service programs are described that are used for basic ser-
vicing work. In addition, the service level provides a series of test pro-
grams with a number of diagnostic and inspection options in the unit.
Test programs are described in the repair instructions and may not be
used without precise knowledge of their function.
No.
Designation
Description
Use
1 – 5
Setup pro-
grams
Repeat of setup programs 1 – 5 from service level 1.
6 –
19
Not assigned.
20
Reset
Front panel reset to
default values
(=default values for
programs, setup and
service settings,
clearing the error
list).
Perform reset:
Press the yellow selection button.
The S (for Service mode) appears
with an asterisk in the top left dis-
play. Now press and hold the top left
selection button until a check mark
appears in the top left display.
21
Error list
Saves up to
20 errors.
Display error list:
Press the yellow selection button.
The error number appears in the
yellow display, the associated error
code in the blue display. The yellow
Plus/Minus buttons can be used to
switch between errors.
Clear error list:
Press and hold the top left selection
button until a check mark in the
adjacent display indicates the end of
the process.
Exit service pro-
gram.
Press the yellow selection button
until only the number of the service
program appears in the yellow dis-
play.
Содержание VIO 100 C
Страница 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Страница 2: ......
Страница 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Страница 6: ...Table of Contents 6 42 80116 221 10 15...
Страница 20: ...4 Service mode 20 42 80116 221 10 15...
Страница 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...