6
•
Maintenance and servicing
38
/ 42
801
16-
221
10.
15
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
1. Set the test specimen to:
DRY CUT, 50 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
Test procedure for VIO 100 C
1. Set the test specimen to:
AUTO CUT, 100 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document the measured value. The tolerance range is
80 to 120 watts.
1. Set the test specimen to:
DRY CUT, 100 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document the measured value. The tolerance range is
80 to 120 watts.
HF power output COAGULATE
Testing and measuring equipment
Erbe Art. No.
Description
20192-127
Patient cable AE
or
Patient cable AE, international
20192-110
20190-045
Electrode handle ICC/ACC
–
Laboratory measuring cable
20194-070
Patient cable NE
or
Patient cable NE, international
20194-075
20100-033
Adapter cable NE