Entek IRD 6652 DSP Protection Monitor
Configuration
ALARMS submenu
The ALARMS submenu has the following options:
ECCNTRY submenu
The options on the ECCNTRY submenu under the User menu work with the options on the
ECCNTRY submenu under the Factory menu (page 30). The options on the two ECCNTRY
submenus define the parameters for vector monitoring. This ECCNTRY submenu has the
following options:
DNG1 HI
ALT1 LO
DNG1 LO
ALT2 HI
DNG2 HI
ALT2 LO
DNG2 LO
EXIT
SET
ALT1 HI
User menu
→
ALARMS
ALT1 HI
Choose the Alert alarm setpoint for input channel 1 (0–9999.99).
DNG1 HI
Choose the Danger alarm setpoint for input channel 1 (0–9999.99).
ALT1 LO
Not used with the 6652 DSP (6682 DSP monitor feature).
DNG1 LO
Not used with the 6652 DSP (6682 DSP monitor feature).
ALT2 HI
Choose the Alert alarm setpoint for input channel 2 (0–9999.99).
DNG2 HI
Choose the Danger alarm setpoint for input channel 2 (0–9999.99).
ALT2 LO
Not used with the 6652 DSP (6682 DSP monitor feature).
DNG2 LO
Not used with the 6652 DSP (6682 DSP monitor feature).
EXIT
Return to the User menu. The monitor remembers the current values for
menu options, but does not write them to memory until you exit the User
menu.
ES ALM2
CAP REF
REF1 A
REF1 P
REF2 A
REF2 P
EXIT
ES ALM1
SET
User menu
→
ECCNTRY
ES ALM1
Choose the vector (eccentricity) Danger alarm setpoint for input channel 1
(0.10–9999.99).
ES ALM2
Choose the vector (eccentricity) Danger alarm setpoint for input channel 2
(0.10–9999.99).
CAP REF
Capture the current amplitude-phase vector to use as the reference:
l
CURR1 - Capture the reference (1x) vector from channel 1 when you
press
SEL.
l
CURR2 - Capture the reference (1x) vector from channel 2 when you
press
SEL.
l
NO - Manually enter channel 1 and 2 reference vector with the REF
options.
REF1 A
Choose the channel 1 reference vector amplitude if CAP REF is NO
(0.00–9999.99).
REF1 P
Choose the channel 1 reference vector phase if CAP REF is NO
(-180.0 to 180.0°).
REF2 A
Choose the channel 2 reference vector amplitude if CAP REF is NO
(0.00–9999.99).
REF2 P
Choose the channel 2 reference vector phase if CAP REF is NO
(-180.0 to 180.0°).
EXIT
Return to the User menu. The monitor remembers the current values for
menu options, but does not write them to memory until you exit the User
menu. The only exception is the CAP REF option, which the monitor sets
when you select CURR 1 or CURR 2.