3 Installation
TopCal S CPC 300
16
EHauser
3.4.3
Water and compressed air connections
Water
3 ... 6 bar,
filtered 100 µm,
max. 56 °C
(Rinse the line
comprehensively before
connecting it!)
!
Note!
Please use commercially available filters
(e.g. of make Grünbeck, Höchstädt a.D.)
Compressed air
4 ... 6 bar (optimum 5
bar)
filtered 0.5 µm, free from
oil and condensate
minimum line diameter
10 mm)
Screw unions
bulkhead gland
OD 6 / ID 4
3.4.4
Chemical
connections
Pump:
(Conveyance of buffer/cleaning agent from
CPG 300 to assembly)
!
Note!
The maximum hose length is 10 m. Of which:
Maximum discharge
height
5 m
Maximum horizontal dis-
charge range
10 m
Pressure resistance of
internal components
up to 12 bar
ƒig. 13:
Maximum horizontal and vertical
installation distances
Suction of buffer/cleaner from the buffer bottles:
Maximum suction height
2 m
"
Caution!
•
Danger of pump damage.
Concentrated acids (e.g. phosphoric acid, nitric acid,
sulfuric acid), acetone, ketones, aromatic solutizing agents as well as hot media
have to be
conveyed
via external additional valves.
• Only convey diluted acids (max. 3%) directly via the pump.
max. 2 m
max. 5 m
C07-CPC300xx-17-12-00-xx-001.EPS
max. 10 m
Hauser
CPG 300