6 Commissioning
Hauser
97
To enter the menu, proceed as follows:
pH calibration
"Manual data entry" (
)
The numeric values for electrode zero point and slope are entered numerically and by
hand.
⇒
⇒
pH
7.00
Cal
Next
Info
E
Hold
pH
7.00
Man. cal. with TopCal/
TopClean only if
service switch active
Cal
Next
Info
E
Hold
CODE
CHOICE
(default = bold)
INFO
C1
Electrode 1
Electrode 2
shared
Abort calibration
Selection for calibration
(only two-circuit)
Select electrode 1
or
2, and then run through
calibration for each individual electrode.
CA
Calibration with
data entry
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
CAA1
025.0 °C
(–20.0 ... +150.0 °C)
Entry of temperature,
(only if "Calibrate with MTC" is selected)
CAA2 / CAA3
07.00
(pH –2.00 ... +16.00)
IsFET:
act.value
(–500 ... +500mV)
Entry of the electrode zero point 1 / 2
Confirm by pressing
F
CAA4 / CAA5
59.16 mV/pH
(5.00 ... 99.00 mV/pH)
Entry of the electrode slope 1 / 2
Confirm by pressing
F
CAA6
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Reject
: The data is not accepted and the device has
not been recalibrated.
Recalibrate
: The data is rejected and the device is
recalibrated.
CAA7
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that measure-
ment can take place?