Technical data
RMA42
44
Hauser
10.0.4
Performance characteristics
Reference operating
conditions
Power supply: 230 V AC, 50/60 Hz
Ambient temperature: 25 °C ± 5 °C (77 °F ± 9 °F)
Humidity: 20 % to 60 % rel. humidity
Maximum measured error
Universal input:
Accuracy
Input:
Range:
Maximum measured error of measuring
range (oMR):
Current
0 to 20 mA, 0 to 5 mA, 4 to 20 mA; overrange: to 22 mA
± 0.05%
Voltage
≥
1 V
0 to 10 V, 2 to 10 V, 0 to 5 V, 1 to 5 V, 0 to 1 V, ± 1 V, ± 10 V, ± 30
V
± 0.1%
Voltage < 1 V
± 100 mV
± 0.05%
Resistance
measurement
30 to 3000
Ω
4-wire: ± (0.10% oMR + 0.8
Ω
)
3-wire: ± (0.10% oMR + 1.6
Ω
)
2-wire: ± (0.10% oMR + 3
Ω
)
Resistance
thermometer
Pt100, -200 to 850 °C (-328 to 1562 °F) (IEC60751,
α
=0.00385)
Pt100, -200 to 850 °C (-328 to 1562 °F) (JIS1604, w=1.391)
Pt100, -200 to 649 °C (-328 to 1200 °F) (GOST,
α
=0.003916)
Pt500, -200 to 850 °C (-328 to 1562 °F) (IEC60751,
α
=0.00385)
Pt1000, -200 to 600 °C (-328 to 1112 °F) (IEC60751,
α
=0.00385)
4-wire: ± (0.10% oMR + 0.3 K (0.54 °F))
3-wire: ± (0.10% oMR + 0.8 K (1.44 °F))
2-wire: ± (0.10% oMR + 1.5 K (2.7 °F))
Cu100, -200 to 200 °C (-328 to 392 °F) (GOST, w=1.428)
Cu50, -200 to 200 °C (-328 to 392 °F) (GOST, w=1.428)
Pt50, -200 to 1100 °C (-328 to 2012 °F) (GOST, w=1.391)
Pt46, -200 to 850 °C (-328 to 1562 °F) (GOST, w=1.391)
Ni100, -60 to 250 °C (-76 to 482 °F) (DIN43760,
α
=0.00617)
Ni1000, -60 to 250 °C (-76 to 482 °F) (DIN43760,
α
=0.00617)
4-wire: ± (0.20% oMR + 0.3 K (0.54 °F))
3-wire: ± (0.20% oMR + 0.8 K (1.44 °F))
2-wire: ± (0.20% oMR + 1.5 K (2.7 °F))
Cu53, -50 to 200 °C (-58 to 392 °F) (GOST, w=1.426)
4-wire: ± (0.30% oMR + 0.3 K (0.54 °F))
3-wire: ± (0.30% oMR + 0.8 K (1.44 °F))
2-wire: ± (0.30% oMR + 1.5 K (2.7 °F))
Thermocouples
Type J (Fe-CuNi), -210 to 1200 °C (-346 to 2192 °F) (IEC60584)
± (0.1% oMR +0.5 K (0.9 °F))
from -100 °C (-148 °F)
Type K (NiCr-Ni), -200 to 1372 °C (-328 to 2502 °F) (IEC60584)
± (0.1% oMR +0.5 K (0.9 °F))
from -130 °C (-202 °F)
Type T (Cu-CuNi), -270 to 400 °C (-454 to 752 °F) (IEC60584)
± (0.1% oMR +0.5 K (0.9 °F))
from -200 °C (-328 °F)
Type N (NiCrSi-NiSi), -270 to 1300 °C (-454 to 2372 °F) (IEC60584) ± (0.1% oMR +0.5 K (0.9 °F))
from -100 °C (-148 °F)
Type L (Fe-CuNi), -200 to 900 °C (-328 to 1652 °F) (DIN43710,
GOST)
± (0.1% oMR +0.5 K (0.9 °F))
from -100 °C (-148 °F)
Type D (W3Re/W25Re), 0 to 2495°C (32 to 4523°F)(ASTME998)
± (0.15% oMR +1.5 K (2.7 °F))
from 500 °C (from 932 °F)
Type C (W5Re/W26Re), 0 to 2320°C (32 to 4208°F)(ASTME998)
± (0.15% oMR +1.5 K (2.7 °F))
from 500 °C (932 °F)
Type B (Pt30Rh-Pt6Rh), 0 to 1820 °C (32 to 3308 °F) (IEC60584)
± (0.15% oMR +1.5 K (2.7 °F))
from 600 °C (1112 °F)
Type S (Pt10Rh-Pt), -50 to 1768 °C (-58 to 3214 °F) (IEC60584)
± (0.15% oMR +3.5 K (6.3 °F))
for -50 to 100 °C (-58 to 212 °F)
± (0.15% oMR +1.5 K (2.7 °F))
for 100 to 1768 °C (212 to 3214 °F)
Type R (Pt13Rh-Pt), -50 to 1768 °C (-58 to 3214 °F) (IEC60584)
± (0.15% oMR +3.5 K (6.3 °F))
for -50 to 100 °C (-58 to 212 °F)
± (0.15% oMR +1.5 K (2.7 °F))
for 100 to 1768 °C (212 to 3214 °F)
Type U (Cu-CuNi), -200 to 600 °C (-328 to 1112 °F) (DIN 43710)
± (0.15% oMR +0.5 K (0.9 °F))
from -100 °C (-148 °F)
Содержание RMA42