Proline Prowirl R 200
Electrical connection
Hauser
39
The device restarts automatically and reports the current address (factory
setting: 126).
2. Set the address via the operating menu: Setup → Communication →Device
address
2
98
3
4
5
6
7
8
1
DIP
ON
A0015903
11
Example of software addressing; switch 8 is set to the "ON" position; the address is defined
in the operating menu (Setup → Device address).
7.4
Ensuring the degree of protection
The measuring device fulfills all the requirements for the IP66/67 degree of protection,
Type 4X enclosure.
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following
steps after the electrical connection:
1. Check that the housing seals are clean and fitted correctly. Dry, clean or replace
the seals if necessary.
2. Tighten all housing screws and screw covers.
3. Firmly tighten the cable glands.
4. To ensure that moisture does not enter the cable entry, route the cable so that it
loops down before the cable entry ("water trap").
A0013960
5. Insert dummy plugs into unused cable entries.
7.5
Post-connection check
Are cables or the device undamaged (visual inspection)?
Do the cables comply with the requirements (→ 24)?
Do the cables have adequate strain relief?