Proline Prowirl R 200
Installation
Hauser
21
6.2.5
Mounting the transmitter of the remote version
!
CAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
‣
Do not exceed the permitted maximum ambient temperature of +50 °C (+122 F).
‣
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly
in warm climatic regions.
!
CAUTION
Excessive force can damage the housing!
‣
Avoid excessive mechanical stress.
Wall mounting
80 (3.15)
80 (3.15)
19 (0.6)
!
8.6 (0.39)
M 8
A0019864
5
Engineering unit mm (in)