Brief Operating Instructions

Proline Prowirl R 200

Vortex flowmeter

These Instructions are Brief Operating Instructions; they do not

replace the Operating Instructions included in the scope of

supply.
For detailed information, refer to the Operating Instructions

and other documentation on the CD-ROM provided or visit

"www.endress.com/deviceviewer".

KA01138D/06/EN/01.13

71216188

Содержание Proline Prowirl R 200

Страница 1: ...Instructions are Brief Operating Instructions they do not replace the Operating Instructions included in the scope of supply For detailed information refer to the Operating Instructions and other doc...

Страница 2: ...s 12 6 2 Mounting the measuring device 19 6 3 Post installation check 23 7 Electrical connection 24 7 1 Connection conditions 24 7 2 Connecting the measuring device 32 7 3 Hardware settings 37 7 4 Ens...

Страница 3: ...rical symbols Symbol Meaning A0011197 Direct current A terminal to which DC voltage is applied or through which direct current flows A0011198 Alternating current A terminal to which alternating voltag...

Страница 4: ...rocedures processes or actions that are preferred A0011184 Forbidden Indicates procedures processes or actions that are forbidden A0011193 Tip Indicates additional information A0011194 Reference to do...

Страница 5: ...epending on the application Following instructions and basic conditions 2 2 Designated use Application and media Depending on the version ordered the measuring device can also measure potentially expl...

Страница 6: ...temperature concentration or level of contamination in the process can alter the corrosion resistance properties Residual risks Possible burn hazard due to fluid temperatures For elevated fluid temper...

Страница 7: ...4 5 6 7 8 9 E A0020649 1 Important components of a measuring device 1 Electronics compartment cover 2 Display module 3 Main electronics module 4 Cable glands 5 Transmitter housing incl HistoROM 6 I O...

Страница 8: ...ance and product identification Proline Prowirl R 200 8 Endress Hauser 4 Incoming acceptance and product identification 4 1 Incoming acceptance A0015502 1 2 1 2 A0013843 A0013695 A0015502 A0013698 A00...

Страница 9: ...numbers from nameplates in W M Device Viewer www endress com deviceviewer All information about the measuring device is displayed For an overview of the scope of the Technical Documentation provided r...

Страница 10: ...g the product WARNING Center of gravity of the measuring device is higher than the suspension points of the webbing slings Risk of injury if the measuring device slips Secure the measuring device from...

Страница 11: ...Proline Prowirl R 200 Storage and transport Endress Hauser 11 A0015604 A0015605...

Страница 12: ...tation Vortex meters require a fully developed flow profile as a prerequisite for correct volume flow measurement The direction of the arrow on the sensor nameplate helps you to install the sensor acc...

Страница 13: ...t 2 Danger of electronics overheating If the fluid temperature is 200 C 392 F orientation B is not permitted for the wafer version Prowirl D with nominal diameters DN 100 4 and DN 150 6 3 In the case...

Страница 14: ...ained at the very minimum If there are several flow disturbances present the longest specified inlet run must be maintained 15 DN 5 DN 1 2 17 DN 8 h 5 DN 7 50 DN 5 DN 6 25 DN 5 DN 40 DN 5 DN 5 4 20 DN...

Страница 15: ...ificate option JN Transmitter ambient temperature 50 C 58 F Remote version Transmitter Non Ex 40 to 80 C 40 to 176 F 1 Ex i 40 to 80 C 40 to 176 F 1 Ex d 40 to 60 C 40 to 140 F 1 ATEX II1 2G Ex d Ex i...

Страница 16: ...radiator and protects the electronics from overheating and excessive cooling NOTICE Electronics overheating on account of thermal insulation Observe the maximum permitted insulation height of the tran...

Страница 17: ...ects on vortex formation in the sensor PT TT 3 5 x DN 4 8 x DN A0019205 PT Pressure measuring point TT Temperature measuring point Open aperture flow conditioner If the required inlet runs cannot be o...

Страница 18: ...the process medium v average flow velocity Installation for delta heat measurements Order code for Sensor version option 3 Mass flow integrated temperature measurement The second temperature measureme...

Страница 19: ...6 2 Mounting the measuring device 6 2 1 Required tools For transmitter For turning the transmitter housing Open ended wrench8 mm For opening the securing clamps Allen key3 mm For sensor For flanges a...

Страница 20: ...ing so that the cable entries do not point upwards A0013964 6 2 4 Mounting the sensor WARNING Danger due to improper process sealing Ensure that the inside diameters of the gaskets are greater than or...

Страница 21: ...ting and housing deformation Do not exceed the permitted maximum ambient temperature of 50 C 122 F If operating outdoors Avoid direct sunlight and exposure to weathering particularly in warm climatic...

Страница 22: ...ng 20 70 0 79 to 2 75 1 4 x SW 13 A0019862 6 Engineering unit mm in 1 Pipe mounting set 6 2 6 Turning the transmitter housing To provide easier access to the connection compartment or display module t...

Страница 23: ...tion document on the CD ROM provided Ambient temperature 15 Measuring range Has the correct orientation for the sensor been selected 12 According to sensor type According to medium temperature Accordi...

Страница 24: ...l regulations Permitted temperature range 40 C 40 F to 80 C 176 F Minimum requirement cable temperature range ambient temperature 20 K Signal cable Current output For 4 20 mA standard installation cab...

Страница 25: ...e with common shield 4 pairs pair stranded and additional steel wire braided sheath Flame resistance According to DIN EN 60332 1 2 Oil resistance According to DIN EN 60811 2 1 Shielding Galvanized cop...

Страница 26: ...nal transmission Input passive supply voltage and signal transmission Ground terminal for cable shield Order code for Output Terminal numbers Output 1 Output 2 Input 1 2 3 4 5 6 Option A 4 20 mA HART...

Страница 27: ...Ex tb and cCSAus Div 1 If a reinforced connecting cable is used The connection is via an M12 connector For all other approvals If the standard connecting cable is used Connection to the connection hou...

Страница 28: ...ne end on the feed side with capacitance termination at the field device Shielding at one end on the feed side Experience shows that the best results with regard to EMC are achieved in most cases in i...

Страница 29: ...Measuring device 6 Local grounding 7 Bus terminator 8 Potential matching line NOTICE In systems without potential matching the multiple grounding of the cable shield causes mains frequency equalizing...

Страница 30: ...ly unit with load 2 The minimum terminal voltage increases if local operation is used see the following table 3 Voltage drop 2 2 to 3 V for 3 59 to 22 mA Increase in minimum terminal voltage Local ope...

Страница 31: ...Ex and Ex d Sample calculation Supply voltage of the supply unit US 19 V Uterm min 12 V measuring device 1 V local operation without lighting 13 V Maximum load RB 19 V 13 V 0 022 A 273 The minimum te...

Страница 32: ...in the device specific Ex documentation 7 2 1 Connecting the transmitter The connection of the transmitter depends on the following order codes Connection version terminals or device plug Connection v...

Страница 33: ...llowing procedure in the action sequence given is recommended for the remote version 1 Mount the transmitter and sensor 2 Connect the connecting cable 3 Connect the transmitter The way the transmitter...

Страница 34: ...ess Hauser Connecting the sensor connection housing 3 mm A0020410 A0020411 1 Wire the connecting cable Terminal 1 brown cable Terminal 2 white cable Terminal 3 yellow cable Terminal 4 green cable 2 Co...

Страница 35: ...e Prowirl R 200 Electrical connection Endress Hauser 35 Connection to the wall holder of the transmitter Connecting the transmitter via plug A0020412 Connecting the transmitter via terminals E 3 mm A0...

Страница 36: ...Electrical connection Proline Prowirl R 200 36 Endress Hauser 8 mm TX 10 A0020405 15 A0020406...

Страница 37: ...onnecting cable Terminal 1 brown cable Terminal 2 white cable Terminal 3 yellow cable Terminal 4 green cable 2 Connect the cable shield via the cable strain relief 7 3 Hardware settings 7 3 1 Setting...

Страница 38: ...d with the software addressing method 2 9 8 3 4 5 6 7 8 1 D IP O N A0015686 9 Address switch in the connection compartment Hardware addressing 1 Set switch 8 to the OFF position 2 Using switches 1 to...

Страница 39: ...degree of protection Type 4X enclosure To guarantee IP66 67 degree of protection Type 4X enclosure carry out the following steps after the electrical connection 1 Check that the housing seals are clea...

Страница 40: ...ect Is the terminal assignment or the pin assignment of the device plug 28 correct If supply voltage is present do values appear on the display module Are all housing covers installed and firmly tight...

Страница 41: ...dividual parts of the operating menu are assigned to certain user roles Each user role corresponds to typical tasks within the device lifecycle For detailed information about the operating philosophy...

Страница 42: ...channel number symbol for diagnostic behavior 1 4 Status area 1 5 Measured value 1 6 Unit for the measured value 1 7 Operating elements 2 Operational display with measured value shown as 1 bar graph...

Страница 43: ...ayed Symbols for measurement channel numbers A0016325 Measurement channel 1 to 4 The measurement channel number is displayed only if more than one channel is present for the same measured variable typ...

Страница 44: ...016620 A0014040 Inserts decimal separator at the input position Inserts minus sign at the input position Clears all entered characters Operating symbols in the text editor A0013985 A0013987 A0013986 C...

Страница 45: ...menu submenu Pressing the key briefly Opens the selected menu submenu or parameter Starts the wizard If help text is open closes the help text of the parameter Pressing the key for 2 s for parameter I...

Страница 46: ...for the selected parameter opens 2 Press simultaneously The help text is closed 8 2 6 User roles and related access authorization The two user roles Operator and Maintenance have different write acce...

Страница 47: ...code appears 2 Enter the access code The symbol in front of the parameters disappears all previously write protected parameters are now re enabled 8 2 8 Enabling and disabling the keypad lock The key...

Страница 48: ...Control system e g PLC 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA195 and Field Communicator 475 4 Field Communicator 475 5 Computer with oper...

Страница 49: ...Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 9 System integration For system integration See Operating Instructions for the device on the CD ROM provided 10 Commission...

Страница 50: ...he device 10 3 Setting the operating language Factory setting English or ordered local language XXXX XXX X X 20 50 Display operat Setup Main menu 0104 1 Language English Espa ol Fran ais Language Engl...

Страница 51: ...er entry Factory setting Device tag Enter tag for measuring point Max 32 characters such as letters numbers or special characters e g Prowirl 10 6 Protecting settings from unauthorized access The foll...

Страница 52: ...operational display mode from the navigation and editing view If write access is activated via access code it can be also be deactivated only via the access code 47 The user role with which the user i...

Страница 53: ...e ON position enables the hardware write protection The Hardware locked option is displayed in the Locking status parameter In addition on the local display the symbol appears in front of the paramete...

Страница 54: ...iagnostics 1 Diagnostics 2 Diagnostics 3 Supply voltage Supply voltage A0013940 EN 14 Message for remedial measures 1 Short text 2 Diagnostic behavior with diagnostic code 3 Service ID 4 Operation tim...

Страница 55: ......

Страница 56: ...www addresses endress com...

Отзывы: