Commissioning
Proline Promag H 100
46
Hauser
10.2
Establishing a connection via FieldCare
• For FieldCare connection (→ 41)
• For establishing a connection via FieldCare (→ 44)
10.3
Configuring the device address via software
Applies only to device version with EtherNet/IP communication type
Software addressing is carried out in the
Communication
submenu.
Navigation path
"Setup" menu → Communication
When delivered, the measuring device has the following factory settings:
Ethernet network and Web server
IP address
192.168.1.212
Subnet mask
255.255.255.0
Default gateway
192.168.1.212
• If hardware addressing is active, software addressing is disabled.
• If a switch is changed to the "ON" position, then the first three octets of the software
address are retained.
If the IP address of the device is not known, the device address currently configured can
be read out: Operating Instructions for the device
10.4
Configuring the measuring device
The
Setup
menu with its submenus is used for fast commissioning of the measuring device.
The submenus contain all the parameters required for configuration, such as parameters for
measurement or communication.
Submenu
Meaning
System units
Configuring the units for all measured values
Medium selection
Defining the medium
Communication
Configuration of the digital communication interface
Low flow cut off
Configuring the low flow cut off
10.5
Defining the tag name
To enable fast identification of the measuring point within the system, you can enter a unique
designation using the
"Device tag" parameter
and thus change the factory setting.