Proline Promag H 100
Installation
Hauser
19
6.2
Mounting the measuring device
6.2.1
Required tools
For sensor
For flanges and other process connections:
• Screws, nuts, seals etc. are not included in the scope of supply and must be provided by the
customer.
• Appropriate mounting tools
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the sensor
!
WARNING
Danger due to improper process sealing!
‣
Ensure that the inside diameters of the gaskets are greater than or equal to that of the
process connections and piping.
‣
Ensure that the gaskets are clean and undamaged.
‣
Install the gaskets correctly.
1. Ensure that the direction of the arrow on the sensor matches the flow direction of the
medium.
2. To ensure compliance with device specifications, install the measuring device between
the pipe flanges in a way that it is centered in the measurement section.
3. Install the measuring device or turn the transmitter housing so that the cable entries do
not point upwards.
A0013964
The sensor is supplied to order, with or without pre-installed process connections. Pre-
installed process connections are firmly secured to the sensor by 4 or 6 hexagonal-headed
bolts.
Содержание Proline Promag H 100 EtherNet/IP
Страница 13: ...Proline Promag H 100 Storage and transport Endress Hauser 13 A0015604 A0015605...