Proline Promag H 100
Commissioning
Hauser
49
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
128
- Write protection
- Default Ethernet
network settings
IP 192.168.1.212
IP
Address setting
(last octet)
A0017915
For device version with PROFIBUS DP communication type
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
- Write protection
- Not used
PROFIBUS
address
- Software addressing
A0021262
‣
Setting the write protection switch on the electronics module to the ON position enables
the hardware write protection.
If hardware write protection is enabled, the
Hardware locked
option is displayed in
the
Locking status
parameter.