Technical data
Liquisys M CCM223/253
104
Hauser
11.2
Output
Output signal
0/4 ... 20 mA, galvanically separated, active
Signal on alarm
2.4 or 22 mA in case of an error
Load
maximum 500
Transmission range
Resolution
max. 700 digits/mA
Isolation voltage
max. 350 V
RMS
/500 V DC
Overvoltage protection
according to EN 61000-4-5
Auxiliary voltage output
HART
Signal coding
Frequency Shift Keying (FSK) + 0.5 mA via current output
signal
Data transfer rate
1200 Baud
Galvanic isolation
yes
PROFIBUS PA
Signal coding
Manchester Bus Powered (MBP)
Data transfer rate
31.25 kBit/s, voltage mode
Galvanic isolation
yes (IO-Module)
PROFIBUS DP
Signal coding
RS485
Data transfer rate
9.6 kBd, 19.2 kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5 MBd
Galvanic isolation
yes (IO-Module)
Cl
2
/ClO
2
:
0 to 10 mg/l for CCS120
0 to 20 mg/l for CCS140/240
0 to 5 mg/l for CCS141/241 and 963
Cl
2
/ClO
2
:
0 to 10 mg/l for OCS120
0 to 20 mg/l for OCS140/240
0 to 5 mg/l for OCS141/241 and 963
Temperature:
0 to 50 °C (32 to 122 °F)
pH:
pH 4 to 9
ORP:
0 to 1500 mV
Output voltage:
15 V ± 0.6
Output current:
max. 10 mA