Liquisys M CCM223/253
Commissioning
Hauser
43
6.4.3
Current input
To use the "Current input" function group, you need a relay board with current input which is
not part of the basic version. With this function group you can monitor process parameters
and use them for feedforward control. For this purpose, you must connect the current output
of an external measured variable (e.g. flowmeter) to the 4 to 20 mA input of the transmitter.
The following assignment applies:
Monitoring of flow in main stream
Flow monitoring in the main stream is particularly practical if the sample flow through the
flow assembly CCA250 is completely independent of the flow in the main stream.
This permits signalling of an alarm condition in the main stream (flow too low or has
completely stopped) and triggers dosing switch-off even if the sample flow is retained due
to the method of installation.
This monitoring method corresponds to monitoring the flow rate in the sample stream (see
SETUP 1).
Flow in main stream
Current signal in mA
Current input signal in
%
Current input lower
range limit
Flowmeter lower setting
value
4
0
Current input upper
range limit
Flowmeter upper setting
value
20
100
a0002019
Fig. 32:
Alarm signalling and dosing switch-off by the main stream
A
B
C
D
E
Flow in main stream
Relay contacts of PID controller
Alarm relay
Flow below switch-off limit Z4 or flow failure
Flow alarm
F
Z2
Z3
0
1
Flow restoration
Delay for controller switch-off, see field Z2
Delay for controller switch-on, see field Z3
Off
On
0
1
1
1
1
0
0
0
A
B
C
D
E
F
G
Z2
Z2
Z3
Z3
Содержание Liquisys M CCM223
Страница 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Страница 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Страница 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Страница 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Страница 124: ...www addresses endress com 71212626...