Liquisys M CCM223/253
Commissioning
Hauser
69
6.4.8
Service
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
S
SERVICE function
group
Service function settings.
S1
Select language
ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
S2
Configure Hold
S+C = Hold during
configuration and calibration
Cal = Hold during calibration
Setup = Hold during
configuration
None = no Hold
S = setup
C = calibration
S3
Manual Hold
Off
On
The setting is retained even in the event of
a power failure.
S4
Enter Hold dwell
period
10 s
0 to 999 s
S5
Enter SW upgrade
release code (Plus
Package)
0000
0000 to 9999
The code is located on the nameplate.
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the ENTER
key.
"1" is displayed if the code is active.
S6
Enter SW upgrade
release code
Chemoclean
0000
0000 to 9999
The code is located on the nameplate.
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the ENTER
key.
"1" is displayed if the code is active.
S7
Order number is
displayed
When the device is upgraded, the order
code is
not
automatically adjusted.
S8
Serial number is
displayed
S
SERVICE
S1
Language
ENG
S2
Auto HOLD
S+C
S3
Man
.
HOLD
off
S4
s
Cont
.
Time
10
S5
0000
PlusCode
S6
CleanCode
0000
S7
EK0005
order
S8
12345678
SerNo
Содержание Liquisys M CCM223
Страница 109: ...Liquisys M CCM223 253 Technical data Endress Hauser 109...
Страница 116: ...Appendix Liquisys M CCM223 253 116 Endress Hauser a0002639 en...
Страница 120: ...Appendix Liquisys M CCM223 253 120 Endress Hauser...
Страница 123: ...Liquisys M CCM223 253 Endress Hauser 123...
Страница 124: ...www addresses endress com 71212626...