Liquistation CSF33
Maintenance
Hauser
5
1.2
Replacing the pump tube
CAUTION
!
Danger of injury due to rotating parts
►
Take the sampler out of service before opening the peristaltic pump.
►
Safeguard the sampler against accidental operation while working on the open peristaltic pump.
Open the peristaltic pump as follows:
1.
Take the sampler out of service by pausing a program that is currently running.
2.
Open the fastening clip (item 3) and push the pump bracket (item 4) upwards.
3.
Remove the knurled head screw (item 7) and open the pump head cover (item 5) to the right.
a0014115
Fig. 1: Opening the peristaltic pump
1
2
3
4
5
6
7
Retainer
Pump tube
Fastening clip
Pump bracket
Pump head cover
Positioning pin
Knurled head screw