Diagnostics menu
Liquistation CSF33
22
Hauser
Current pump run time
Read only
Power supply
Read only
The current supply voltage is displayed.
24 V ±0.5 V
Current
Read only
The current consumption of the pump is displayed.
Vacuum
Read only
The vacuum is an indicator of the suction height.
-> 100 mbar is equivalent to a suction height of approx.
1 m.
Probe detected
Read only
Yes: the medium was detected
No: no medium was detected
Vacuum pump
(only for version with vacuum
pump)
Action
Bottle configuration
Read only
Bottle volume
Read only
Distributor position
Options
• Front
• Bottle 1
...
• Back
Select which bottle should be filled with the sample.
Sample volume
Factory setting
200 ml
The sample volume is preset at the factory.
Start sampling
Action
Perform sampling manually.
Action message
Read only
The progress of the sampling operation is displayed.
Power supply
Read only
The current supply voltage is displayed.
24 V ±0.5 V
Current
Read only
The current consumption of the pump is displayed.
Medium LF1
Read only
• Medium detection conductivity 1 switchoff
• Medium detection conductivity 2 disconnection from
protective circuit
-> Both "No" at the start
-> If "Yes", clean conductivity 2
Medium LF2
Read only
Distribution arm
Action
Only for bottle configurations with more than one bottle.
Test distribution arm
Read only
When the menu item is activated, the distribution arm
undergoes a test run. Afterwards, the system moves to
each position in succession and the position is displayed.
In the case of plate distribution, the arm moves left and
right to ensure the bottles are numbered consecutively.
Calibrate the distribution arm if the arm is not
positioned precisely over the bottles.
Position
Read only
Diagnostics/Systemtest/Reset
Function
Options
Info