Troubleshooting
Liquistation CSF33
28
Hauser
262
Module connect.
F
On
On
Electronics module not communicating
1.
Check cable connection, replace if
necessary
2.
Check the power supply of the sampling
control module
3.
Contact the Service Department
263
Electr. module
F
On
On
Wrong kind of electronics module
1.
Replace the module
2.
Contact the Service Department
284
Firmware update
M
On
Off
Update completed successfully
285
Update error
F
On
On
Firmware update failed
1.
Repeat update
2.
SD card error --> use another card
3.
Incorrect firmware --> repeat with suitable
firmware
4.
Contact the Service Department
302
Battery low
M
On
Off
Buffer battery of real time clock is low
The date and time are lost if the power is
interrupted.
--> Contact the Service Department (battery
replacement)
304
Module data
F
On
On
At least 1 module has incorrect configuration
data
1.
Check the system information
2.
Contact the Service Department
305
Power consumption
F
On
On
Total power consumption too high
1.
Check installation
2.
Remove sensors/modules
306
Software
F
On
On
Internal firmware error
--> Contact the Service Department
310
Temp. sensor
F
On
On
Sensor for climate control module
1.
Check the sensor has been fitted correctly
2.
Replace sensor
311
Temp. sensor
F
On
On
Sensor for sample temperature regulation
1.
Check the sensor has been fitted correctly
2.
Replace sensor
312
Temp. sensor
F
On
On
Sensor for outside temperature
1.
Check the sensor has been fitted correctly
2.
Outside temperature inside specified range
(not > - 30
˚
C)
3.
Replace sensor
313
Liquidsensor 1
F
On
On
Contact electrodes fouled
Reliable sample detection not possible
1.
Clean the sensors
2.
Contact the Service Department
No.
Message
Factory settings
Tests or remedial measures
Cat.
Diag.
on/off
Error
current