Liquistation CSF33
Troubleshooting
Hauser
29
314
No sample
F
On
On
No inflow of sample
--> Check pump tube for leaks
315
Refrigeration
F
On
On
Refrigeration regulation not possible
--> Contact the Service Department
316
Heating
F
On
On
Heating regulation not possible
--> Contact the Service Department
317
Sensor
F
On
On
Sensor conductivity 1 fouled
--> Clean sensor
318
Sensor
F
On
On
Sensor conductivity 1 defective
--> Replace sensor
319
Sensor
F
On
On
Sensor conductivity 2 fouled
--> Clean sensor
320
Sensor
F
On
On
Sensor conductivity 2 defective
--> Replace sensor
321
Sensor
F
On
On
Capacitance sensor defective
--> Replace sensor
322
Read subprogram
F
On
On
Selected subprogram cannot be read from the
program memory
--> Create new subprogram
323
Write subprg.
F
On
On
Subprogram created cannot be saved
--> Reset the software
324
Delete subprg.
F
On
On
Selected subprogram cannot be deleted from
the program memory
--> Reset the software
325
Read subprg. list
F
On
On
Subprogram list cannot be read from the pro-
gram memory
--> Reset the software
326
Membrane pump
F
On
On
Vacuum pump defective
--> Contact the Service Department
327
Air-Manager
F
On
On
Runtime error has occurred
• Photoelectric barrier defective
• Cable defective
--> Contact the Service Department
328
Distribution arm
F
On
On
Distribution arm zero point not found
1.
Perform distribution arm test
2.
Contact the Service Department
329
Pump error
F
On
On
Vacuum pump defective
--> Contact the Service Department
330
Membrane pump
F
On
On
Vacuum pump control defective
--> Contact the Service Department
No.
Message
Factory settings
Tests or remedial measures
Cat.
Diag.
on/off
Error
current