Liquistation CSF33
Troubleshooting
Hauser
33
3.3
Device-specific errors
972
Current > 20 mA
M
On
Off
Current output range exceeded
973
Current < 4 mA
M
On
Off
Current output range undershot
974
Diag. confirmed
C
Off
Off
User has acknowledged the message displayed
in the measuring menu.
974
Diag. confirmed
C
Off
Off
User has acknowledged the message displayed
in the measuring menu.
990
Deviation limit
F
On
On
Redundancy: limit value of percentage deviation
exceeded
991
CO2 conc. range
F
On
On
CO
2
concentration (degassed conductivity)
outside the measuring range
992
pH calc range
F
On
On
pH calculation outside the measuring range
993
rH calc range
F
On
On
rH calculation outside the measuring range
994
Dual cond range
F
On
On
Dual conductivity outside the measuring range
Problem
Possible cause
Tests and/or remedial measures
The device cannot be switched
on/display remains dark
No supply voltage
Check if voltage supplied
Basic module defective
Replace basic module
Values appear on display but:
– Display does not change
and / or
– Device cannot be operated
Module not wired correctly
Check modules and wiring
Impermissible operating system condition
Switch off device and switch it on again
Control signals are not
accepted or outputs do not
switch
Incorrect program setting
Check program setting
Incorrect wiring
Check wiring
Electronics failure
Replace basic module
No.
Message
Factory settings
Tests or remedial measures
Cat.
Diag.
on/off
Error
current