Diagnostics menu
Liquistation CSF33
16
Hauser
Go to date
User input
• Go to date
• Time
Use this function to go directly to a specific time in the list. In
this way, you avoid having to scroll through all the
information. The complete list is always visible, however.
Delete all entries
Action
You can delete all the calibration logbook entries here.
Save logbooks
Options
• All event logbooks
• Calibration logbook
• Diagnostic logbook
• Configuration logbook
• HW version logbook
• Version logbook
Use this function to save the logbook to an SD card. You can
then open the file you saved (.csv) at the PC and process it in
MS-Excel for example.
Insert the SD card into the controller card reader and select
the logbook to be saved.
File format
Options
• CSV
• FDM
Save the logbook in the preferred file format.
You can then open the CSV file you saved on the PC in MS
Excel, for example, and process it further
1)
.
You can import the FDM files into Fieldcare and archive
them so they are tamper-proof.
The file name is made up of the "Logbook ident" (Menu/Setup/General settings/Logbooks), an abbreviation for the
particular logbook and a time stamp.
1)
CSV files use international number formats and separators. Therefore they must be imported into MS Excel as external data
with the correct format settings. If you double-click the file to open it, the data are only displayed correctly if MS Excel is
installed with the US country setting.
Diagnostics/Logbooks
Function
Options
Info
Содержание Liquistation CSF33
Страница 41: ...Liquistation CSF33 Spare parts Endress Hauser 41...
Страница 43: ...Liquistation CSF33 Endress Hauser...
Страница 44: ...BA488C 07 EN 14 11 71134632 FM...