Diagnostics menu
Liquistation CSF33
24
Hauser
Diagnostics/Simulation
Function
Options
Info
Current output x:y
Simulation of an output current
This menu appears once for each current output.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the value at the current output, this is
indicated on the display by a simulation icon in front of the
current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the relay state, this is indicated on the display
by a simulation icon in front of the relay display.
State
Options
• Low
• High
Factory setting
Low
Set the desired state.
The relay switches in accordance with your setting when
you switch on the simulation. The display shows "On"
(="Low") or "Off" (="High") for the simulated relay state.
Meas. inputs
Simulation of a measured value
This menu appears once for each measuring input.
Channel : parameter
Sim. main value
Options
• On
• Off
Factory setting
Off
If you simulate the measured value, this is indicated on the
display by a simulation icon in front of the measured value.
Main value
Depends on the sensor
Factory setting
Depends on the sensor
Set the desired simulation value.
Sim. temperature
Options
• On
• Off
Factory setting
Off
If you simulate the temperature measured value, this is
indicated on the display by a simulation icon in front of the
temperature.
Temp. value
-50.0 to +250.0
˚
C
(-58.0 to 482.0
˚
F)
Factory setting
20.0
˚
C (68.0
˚
F)
Set the desired simulation value.