Endress+Hauser Liquistation CSF33 Скачать руководство пользователя страница 1

Liquistation CSF33
BA488C/07/EN/14.11
71134632

Valid as of:
Software version 01.02.00

Operating Instructions

Liquistation CSF33

Automatic sampler for liquid media

Maintenance & diagnostics

Содержание Liquistation CSF33

Страница 1: ...Liquistation CSF33 BA488C 07 EN 14 11 71134632 Valid as of Software version 01 02 00 Operating Instructions Liquistation CSF33 Automatic sampler for liquid media Maintenance diagnostics...

Страница 2: ...Accessories and spare parts This manual does not include the following Setup General settings Operating Instructions BA479C Commissioning Display Operation Operating Instructions BA479C Commissioning...

Страница 3: ...on 19 2 4 Status of inputs outputs 20 2 5 Systemtest Reset 20 2 6 Term information 23 2 7 Simulation 23 3 Troubleshooting 25 3 1 Troubleshooting 25 3 2 System error messages 25 3 3 Device specific err...

Страница 4: ...ESD 1 1 Recommended maintenance Maintenance work has to be carried out at regular intervals to ensure the efficient operation of the sampler The maintenance work comprises Replacing the wear parts Cl...

Страница 5: ...he open peristaltic pump Open the peristaltic pump as follows 1 Take the sampler out of service by pausing a program that is currently running 2 Open the fastening clip item 3 and push the pump bracke...

Страница 6: ...and insert the marking ring into the groove item 9 see 2 7 Close the pump head cover and screw it tight Close the pump bracket 8 Under Menu Diagnostics Term information Pump tube life reset the tube l...

Страница 7: ...ng agents NOTICE Prohibited cleaning agents Damage to the housing surface or housing seal For cleaning purposes never use concentrated mineral acids or bases Never use organic cleaners such as benzyl...

Страница 8: ...the wetted parts as follows 1 Release the thread adapter nut on the intake hose item 6 2 Turn the intake hose to the open position at the hose lock item 5 and remove the hose from above 3 Release the...

Страница 9: ...clear water to the tube connection 3 Remove the bottles from the sample compartment 4 Rinse the wetted parts with clear water by taking a manual sample or by performing a pump test under Menu Diagnos...

Страница 10: ...t of service by pausing a program that is currently running 2 Open the peristaltic pump as described in the Replacing the pump tube section 3 Remove the pump tube 4 Remove any silicone deposits on the...

Страница 11: ...s in reverse order 1 3 3 Sample compartment The sample compartment has a fully integrated plastic inner lining Clean the sample compartment as follows 1 Remove the bottle trays and the distribution pa...

Страница 12: ...information on calibration is provided in BA489C Calibration The position of the distribution arm is set at the factory The sample volume of the peristaltic pump is also calibrated at the factory The...

Страница 13: ...ling are deleted automatically with the next successful sampling Diagnostics messages caused by the level of medium in the bottle are deleted the next time the program is started If diagnostics messag...

Страница 14: ...Yes Debugging logbook Only accessible with the special activation code Service 1000 Yes No Yes Yes Diagnostics Logbooks Function Options Info Program logbook Chronological list of the programming even...

Страница 15: ...ist of the configuration events Show Events are displayed Select a particular event to display more detailed information Go to date User input Go to date Time Use this function to go directly to a spe...

Страница 16: ...mple Insert the SD card into the controller card reader and select the logbook to be saved File format Options CSV FDM Save the logbook in the preferred file format You can then open the CSV file you...

Страница 17: ...t BasicSampling or PrgStop the name of the program appears In the case of StdSampling PrgPartStart or PrgPartStop the name of the subprogram appears Bottle configuration 12x 6x PE glass plate distribu...

Страница 18: ...me after date time Event when an event occurs Stop date 06 05 2010 Field only populated if Program end Date Time The time the program was stopped is displayed Start flow sum unit 100 m Field only popu...

Страница 19: ...to the hardware and you can enter the new code here Serial number Read only The serial number allows you to access device data and documentation on the Internet www products endress com device viewer...

Страница 20: ...upply Read only The current supply voltage is displayed 24 V 0 5 V Overcurrent Read only No no error Yes the fan in the climate control module is defective Contact the Service Department Temp sample r...

Страница 21: ...change the sample volume in the version with the peristaltic pump Sample volume Factory setting 200 ml The sample volume is preset at the factory in the version with the vacuum pump Start sampling Ac...

Страница 22: ...Action message Read only The progress of the sampling operation is displayed Power supply Read only The current supply voltage is displayed 24 V 0 5 V Current Read only The current consumption of the...

Страница 23: ...version with peristaltic pump Number of all samples taken and sample errors Pump tube life for version with peristaltic pump Displays how old the tube is in days hours and minutes This counter must be...

Страница 24: ...display State Options Low High Factory setting Low Set the desired state The relay switches in accordance with your setting when you switch on the simulation The display shows On Low or Off High for t...

Страница 25: ...rent diagnostics messages displayed is provided in the DIAG Diagnostics list menu In accordance with Namur specification NE 107 the diagnostics messages are characterized by Message number Error categ...

Страница 26: ...o you want to change the error category c Should an error current be output d Do you want to trigger a cleaning program 4 Deactivate the message for example Diagnostics message to Off Configuration op...

Страница 27: ...rigger a cleaning program You can define the cleaning programs under Menu Setup Additional functions Cleaning Detail information Read only Here you can find more information on the diagnostic message...

Страница 28: ...nt 304 Module data F On On At least 1 module has incorrect configuration data 1 Check the system information 2 Contact the Service Department 305 Power consumption F On On Total power consumption too...

Страница 29: ...Create new subprogram 323 Write subprg F On On Subprogram created cannot be saved Reset the software 324 Delete subprg F On On Selected subprogram cannot be deleted from the program memory Reset the...

Страница 30: ...m diagnostic settings are displayed Change the pump tubing 339 Sensor M On Off Sensor conductivity 1 fouled Default values from diagnostic settings are displayed Clean sensor 340 Sensor M On Off Senso...

Страница 31: ...354 Bottle check M On Off No empty bottles available for current program No further sampling 355 Start time over M On Off Start time entered is in the past Enter a new start time 356 Overfill check S...

Страница 32: ...easures 1 Check sensor installation 2 Clean sensor 3 Adjust assignment of current outputs 461 Current overrun F On On 502 No text catalog F On On Contact the Service Department 503 Language change M O...

Страница 33: ...range 993 rH calc range F On On rH calculation outside the measuring range 994 Dual cond range F On On Dual conductivity outside the measuring range Problem Possible cause Tests and or remedial measur...

Страница 34: ...rect bottle filled Incorrect distribution selected in the controls No sample cooling Check the setting for the sample chamber temperature in the controls Cooling system defective Have repaired by E H...

Страница 35: ...iter 0 26 US gal PE cap Distributor plate Locating insert 71111158 Distributor plate for 2 x 6 bottles 71111159 Distributor plate for 2 x 12 bottles Bottles caps 71111164 1 liter 0 26 US gal PE cap 24...

Страница 36: ...with peristaltic pump Item No Designation and contents Order number Spare parts kit 13 Roller 71103288 14 Connecting tube with connection 71110972 15 Feedthrough sample infeed ID 10 with cylinder pin...

Страница 37: ...F33 Spare parts Endress Hauser 37 5 2 Vacuum pump a0013811 Fig 9 Spare parts for version with vacuum pump view in front of pump holder a0013812 Fig 10 Spare parts for version with vacuum pump view beh...

Страница 38: ...O ring 71103168 Plastic dosing chamber 350 ml with O ring 71103173 6 Conductive dosing chamber flange with conductivity sensors insulation sleeves and O ring 71102985 7 Capacitance dosing chamber fla...

Страница 39: ...3 Climate control module a0013808 Fig 11 Spare parts view of climate control module Item No Designation and contents Order number Spare parts kit 21 Complete climate control module unpainted 71092603...

Страница 40: ...gnation and contents Order number Spare parts kit 31 Complete distribution arm drive shaft 71113519 32 Distribution arm motor with housing and securing screws 71101959 33 Gasket for dosing chamber doo...

Страница 41: ...Liquistation CSF33 Spare parts Endress Hauser 41...

Страница 42: ...ific diagnostics messages 27 Diagnostic list 13 Error messages 25 Logbooks 14 Measured values 20 Message classification 25 Simulation 23 System information 19 Term information 23 E Error categories 25...

Страница 43: ...Liquistation CSF33 Endress Hauser...

Страница 44: ...BA488C 07 EN 14 11 71134632 FM...