Levelflex M FMP40 with HART/4...20 mA
Troubleshooting
Hauser
71
The measured values
for the interface and the
total level are identical
Echo threshold for the
total level too high
Yes
→
Check the DK value setting in
"Medium Properties" (003).
If the interface layers
are thin, the total level
jumps to the interface
level
The thickness of the
upper phase is less than
60 mm.
Yes
→
Interface measurement is only possible
if the thickness of the interface is
greater than 60 mm.
Interface measured
value jumps
Emulsion layer present
Yes
→
Emulsion layer affects the measurement.
Please contact Hauser.
F m/ft
E m/ft
t
→
actual
expected
F m/ft
E m/ft
t
→
actual
expected
F m/ft
E m/ft
t
→
actual
actual
expected
expected